(a) CardinalHealth Salem Sump Dual Lumen Stomach Tube with Anti-Reflux Valve, 12 Fr/Ch (4.0 mm) x 48 in. (122 cm), Model

Cardinal Health 200, LLC

Summary

The FDA issued a Class II for (a) CardinalHealth Salem Sump Dual Lumen Stomach Tube with Anti-Reflux Valve, 12 by Cardinal Health 200, LLC. Reason: The firm received reports of breakage in the Anti-Reflux Valve (ARV) due to improper use. As a result, changes to labeling were made..

Details

Source

Device Recall

External ID

Z-3302-2024

Action Date

2024-10-02

Status

Ongoing

Category

device

Product Description

(a) CardinalHealth Salem Sump Dual Lumen Stomach Tube with Anti-Reflux Valve, 12 Fr/Ch (4.0 mm) x 48 in. (122 cm), Model #8888266114, Sterile; (b) CardinalHealth Salem Sump Dual Lumen Stomach Tube with Anti-Reflux Valve, 14 Fr/Ch (4.7 mm) x 48 in. (122 cm), Model #8888266122, Sterile; (c) CardinalHealth Salem Sump Dual Lumen Stomach Tube with Anti-Reflux Valve, 16 Fr/Ch (5.3 mm) x 48 in. (122 cm), Model #8888266130, Sterile; and (d) CardinalHealth Salem Sump Dual Lumen Stomach Tube with Anti-Reflux Valve, 18 Fr/Ch (6.0 mm) x 48 in. (122 cm), Model #8888266148, Sterile.

Lot/Code Info: All lot numbers. UDI-DI numbers: (a) 8888266114, UDI-DI 10192253012781; (b) 8888266122, UDI-DI 10192253012804; (c) 8888266130, UDI-DI 10192253012828; and (d) 8888266148, UDI-DI 10192253012842.

Quantity Affected: 24,761,100 tubes total

Reason for Recall

The firm received reports of breakage in the Anti-Reflux Valve (ARV) due to improper use. As a result, changes to labeling were made.

Distribution

Distribution was nationwide, including Puerto Rico and Guam. There was also government/military distribution. Foreign distribution was made to Chile.

Type: Voluntary: Firm initiated

Recall Initiated: 2024-07-24

Company

Context & Analysis

This is a Class II recall. Class II recalls indicate that use of or exposure to the product may cause temporary or medically reversible health consequences, or that the probability of serious consequences is remote.

This was one of 244 device recalls issued in the same week, part of 413 device-related FDA actions this month.

Cardinal Health 200, LLC has 153 FDA actions in our database, including 152 recalls and 1 clearance.

Frequently Asked Questions

What is a Class II recall?

A Class II recall means the product may cause temporary or medically reversible adverse health consequences, or the probability of serious adverse consequences is remote. Consumers should discontinue use.

What should I do if I have this product?

Stop using the product. Check the product against the specific lot numbers or identifiers listed in the recall. Contact the manufacturer (Cardinal Health 200, LLC) or your retailer for a refund or replacement. If you experienced adverse effects, report them to the FDA via MedWatch.

How often does Cardinal Health 200, LLC have FDA actions?

Cardinal Health 200, LLC has 153 FDA actions in our database, including 152 recalls and 1 device clearance.

Where can I find the official FDA record?

The official record is available on FDA.gov. Search for "Z-3302-2024" on FDA.gov for the full record. You can also use the openFDA API to query this record programmatically.

Want to know the moment something like this happens?

Get alerts for device

Related Actions