RecallHawk
Class II Recall

Brand Name: VERIQA; Product Name: RT MonteCarlo 3D v3; Model/Catalog Number: S070031.520; Software Version: 3.1.18;

PTW FREIBURG

Summary

The FDA issued a Class II for Brand Name: VERIQA; Product Name: RT MonteCarlo 3D v3; Model/Catalog Number: S by PTW FREIBURG. Reason: When a bolus is included in a treatment plan and sent for secondary dose calculation, the software accounts for it but does not display this clearly, .

Details

Source

Device Recall

External ID

Z-3174-2026

Action Date

2026-09-23

Status

Ongoing

Category

device

Product Description

Brand Name: VERIQA; Product Name: RT MonteCarlo 3D v3; Model/Catalog Number: S070031.520; Software Version: 3.1.18;

Lot/Code Info: Model/Catalog Number: S070031.520; UDI-DI: +EPTWS070031Q5200; Software Version 3.1.18;

Quantity Affected: 2

Reason for Recall

When a bolus is included in a treatment plan and sent for secondary dose calculation, the software accounts for it but does not display this clearly, which may lead users to incorrectly omit the bolus during treatment and deliver the prescribed dose to the wrong tissue depth.

Distribution

Worldwide - US Nationwide distribution in the states of GA, NY and the countries of Austria, Belgium, Switzerland, Czechia, Germany, Spain, France, Netherlands, Norway, Poland, Portugal, Romania, T¿rkiye, Chile, United Arab Emirates, Saudi Arabia, Mexico, Hong Kong, Brazil, Tajikistan, Ecuador, Iran, Thailand, Serbia, South Africa, Kazakhstan, Algeria.

Type: Voluntary: Firm initiated

Recall Initiated: 2026-08-04

Company

PTW FREIBURG

Freiburg Im Breisgau, N/A

Context & Analysis

This is a Class II recall. Class II recalls indicate that use of or exposure to the product may cause temporary or medically reversible health consequences, or that the probability of serious consequences is remote.

This was one of 203 device recalls issued in the same week, part of 445 device-related FDA actions this month.

PTW FREIBURG has 3 FDA actions in our database, including 3 recalls.

Frequently Asked Questions

What is a Class II recall?

A Class II recall means the product may cause temporary or medically reversible adverse health consequences, or the probability of serious adverse consequences is remote. Consumers should discontinue use.

What should I do if I have this product?

Stop using the product. Check the product against the specific lot numbers or identifiers listed in the recall. Contact the manufacturer (PTW FREIBURG) or your retailer for a refund or replacement. If you experienced adverse effects, report them to the FDA via MedWatch.

How often does PTW FREIBURG have FDA actions?

PTW FREIBURG has 3 FDA actions in our database, including 3 recalls and 0 device clearances.

Where can I find the official FDA record?

The official record is available on FDA.gov. Search for "Z-3174-2026" on FDA.gov for the full record. You can also use the openFDA API to query this record programmatically.

Want to know the moment something like this happens?

Get alerts for device

Related Actions