RecallHawk
Class I Recall

Ivenix Infusion System (IIS), LVP Software; LVP-SW-0004, Software Version 5.9.1 and prior. Infusion pump software for i

Fresenius Kabi USA, LLC

Summary

The FDA issued a Class I for Ivenix Infusion System (IIS), LVP Software; LVP-SW-0004, Software Version 5.9.1 by Fresenius Kabi USA, LLC. Reason: The software has anomalies that have the potential to cause alarms, nonfunctioning pump, or unresponsive screen while continuing therapy. These could.

Details

Source

Device Recall

External ID

Z-3153-2024

Action Date

2024-10-02

Status

Ongoing

Category

device

Product Description

Ivenix Infusion System (IIS), LVP Software; LVP-SW-0004, Software Version 5.9.1 and prior. Infusion pump software for infusion management system.

Lot/Code Info: UDI-DI: 00811505030122; Software Version 5.9.1 and prior

Quantity Affected: 17 units

Reason for Recall

The software has anomalies that have the potential to cause alarms, nonfunctioning pump, or unresponsive screen while continuing therapy. These could cause serious patient harm or death.

Distribution

US Nationwide.

Type: Voluntary: Firm initiated

Recall Initiated: 2024-08-30

Company

Fresenius Kabi USA, LLC

North Andover, MA

Context & Analysis

This is a Class I recall. Class I recalls are the most serious — there is a reasonable probability that use of or exposure to the product will cause serious health consequences or death.

This was one of 244 device recalls issued in the same week, part of 403 device-related FDA actions this month.

Fresenius Kabi USA, LLC has 46 FDA actions in our database, including 46 recalls.

Frequently Asked Questions

What is a Class I recall?

A Class I recall is the most serious type. The FDA has determined there is a reasonable probability that the product will cause serious adverse health consequences or death. Consumers should stop using the product immediately.

What should I do if I have this product?

Stop using the product. Check the product against the specific lot numbers or identifiers listed in the recall. Contact the manufacturer (Fresenius Kabi USA, LLC) or your retailer for a refund or replacement. If you experienced adverse effects, report them to the FDA via MedWatch.

How often does Fresenius Kabi USA, LLC have FDA actions?

Fresenius Kabi USA, LLC has 46 FDA actions in our database, including 46 recalls and 0 device clearances.

Where can I find the official FDA record?

The official record is available on FDA.gov. Search for "Z-3153-2024" on FDA.gov for the full record. You can also use the openFDA API to query this record programmatically.

Want to know the moment something like this happens?

Get alerts for device

Related Actions