Seer Home System is an ambulatory electroencephalograph (EEG) system intended for at-home use for extended EEG and video
Summary
The FDA issued a Class II for Seer Home System is an ambulatory electroencephalograph (EEG) system intended fo by SEER MEDICAL PTY LTD. Reason: During CAPA investigations of sight production inconsistencies, it was highlighted that previous Electromagnetic Compatibility (EMC) assessments relat.
Details
Source
Device Recall
External ID
Z-3143-2024
Action Date
2024-09-25
Status
Ongoing
Category
device
Product Description
Seer Home System is an ambulatory electroencephalograph (EEG) system intended for at-home use for extended EEG and video EEG (vEEG) recordings and analysis. The Seer Home System includes the Sight Monitoring Hub, the Sense wearable, Monitoring Portal and associated accessory components (Disposable EEG Wires, WaterTabs, WaterTabs Application Kit, and Take-home Kit).
Lot/Code Info: 1. Product Number: SM-03-8001-INT; UDI-DI: 09355751001037; Serial Numbers: 000089, 000090, 000091, 000092, 000095, 000182, 000201, 000207, 000208, 000211, 000225, 000245, 000258, 000259, 000285, 000286, 000289, 000290, 000291, 000292, 000296, 000297, 000298, 000299, 000301, 000305, 000307, 000308, 000309, 000310, 000312, 000313, 000316, 000321, 000322, 000323, 000094, 000183, 00028, 000302, 000303, 000306, 000311. 2. Product Number: SM-01-8001; UDI-DI: 09355751001020; Serial Numbers: 000074, 000084, 000113, 000118, 000119, 000187, 000188, 000191, 000223, 000224, 000225, 000226, 000228, 000229, 000230, 000232, 000239, 000241, 000242, 000243, 000244, 000245, 000247, 000250, 000251, 000259, 000262, 000263, 000277, 000279, 000280, 000282, 000283, 000286, 000287, 000288, 000289, 000290, 000291, 000292, 000294, 000297, 000299, 000300, 000301, 000302, 000303, 000304, 000305, 000306, 000307, 000308, 000309, 000310, 000312, 000313, 000315, 000317, 000318, 000319, 000320, 000321, 000322, 000323, 000324, 000327, 000328, 000329, 000333, 000335, 000336, 000337, 000339, 000340, 000341. 3. Product Number: Monitoring Portal; UDI-DI: 09355751001136; Serial Numbers: DMPD80RCMDFX, DMPFRRE7Q1GJ, DMPFT3PDQ1KQ, DMPFT43HQ1GG, DMPFT46JQ1KQ, DMPFT4CYQ1KQ, DMPFT4X1Q1KQ, DMPFV0HGQ1KQ, DMPFV178Q1KQ, DMPFV19SQ1KQ, DMPFV1AFQ1KQ, DMPFV1AXQ1KQ, DMPFV1DBQ1KQ, DMPFV1KQQ1KQ, DMPFV1L2Q1KQ, DMPFV1LNQ1KQ, DMPFV1PRQ1KQ, DMPFV1Q8Q1KQ, DMPFV1R2Q1KQ, DMPFV1R9Q1KQ, DMPFV1SRQ1KQ, DMPFV1TEQ1KQ, DMPFV1WYQ1KQ, DMPFV1YGQ1KQ, DMPFV1ZBQ1KQ, DMPFV1ZMQ1KQ, DMPFV20DQ1KQ, DMPFV20PQ1KQ, DMPFV219Q1KQ, DMPFV21GQ1KQ, F9FF60LYQ1KQ, F9FF70Z5Q1KQ, F9FF71PAQ1KQ, F9FFQ413Q1GJ, F9FFQC4WQ1GG, F9FFQF9VQ1GG. 4. Product Number: SM-01-9005; UDI-DI: 09355751001112; Serial Number: 006240129. 5. Product Number: SM-07-8001; UDI-DI: 09355751001129; Serial Numbers: 007230809, 007230810, 007230823, 007230825, SM-07-9001_230829, 007231004, 007231005, 007231012, 007231023, 007231025, 007231102, 007231108, 007231109, 007231115, 007231120, 007231124, 00720240129, 00720240205, 00720240207, 00720240314, 007240318, 007240319, 007240327, 007240325, 007240326, 007240402, 007240403, 007240507, 007240515, 007240516, 007240522, 007240527, 007240531, 007240612, 007240613, 007240624, 007240626, 007240627, 007240628, 007240702, 007240703. 6. Product Number: SM-01-9003-US; UDI-DI: 09355751001099; Serial Numbers: 230525, 004230609, 004230913, 004231121, 004240306, 004240403, 004240424, 004240502, 004240509, 004240531, 004240603, 004240626, 004240705. 7. Product Number: SM-01-9008; UDI-DI: 09355751001105; Serial Numbers: 00223058, 00223059, 00223066, 00224008, 00224011, 00224013, 00224017, 00224018, 00224021.
Quantity Affected: 3,948 units
Reason for Recall
During CAPA investigations of sight production inconsistencies, it was highlighted that previous Electromagnetic Compatibility (EMC) assessments related to RF radiated emissions may not be reasonably equivalent to the current product due to design changes in the system. A new CAPA was opened to investigate the radiated emissions compliance issues. - Seer Home Systems distributed and retained in the field were not compliant with Electromagnetic Compatibility (EMC) testing according to IEC 60601-1-2 - Seer Home Systems, including accessories, were manufactured and distributed without application of proper design change and manufacturing controls - Seer Home Systems were distributed with labeling deficiencies
Distribution
Worldwide - US Nationwide distribution in the states of MN, NY, PA and the countries of Australia.
Type: Voluntary: Firm initiated
Recall Initiated: 2024-08-12
Company
Melbourne
Context & Analysis
This is a Class II recall. Class II recalls indicate that use of or exposure to the product may cause temporary or medically reversible health consequences, or that the probability of serious consequences is remote.
This was one of 257 device recalls issued in the same week, part of 403 device-related FDA actions this month.
SEER MEDICAL PTY LTD has 2 FDA actions in our database, including 1 recall and 1 clearance.
Frequently Asked Questions
What is a Class II recall?
A Class II recall means the product may cause temporary or medically reversible adverse health consequences, or the probability of serious adverse consequences is remote. Consumers should discontinue use.
What should I do if I have this product?
Stop using the product. Check the product against the specific lot numbers or identifiers listed in the recall. Contact the manufacturer (SEER MEDICAL PTY LTD) or your retailer for a refund or replacement. If you experienced adverse effects, report them to the FDA via MedWatch.
How often does SEER MEDICAL PTY LTD have FDA actions?
SEER MEDICAL PTY LTD has 2 FDA actions in our database, including 1 recall and 1 device clearance.
Where can I find the official FDA record?
The official record is available on FDA.gov. Search for "Z-3143-2024" on FDA.gov for the full record. You can also use the openFDA API to query this record programmatically.
Want to know the moment something like this happens?
Get alerts for deviceRelated Actions
MAHURKAR 12 Fr High Pressure Triple Lumen Acute Dialysis Catheter, 24 cm, Straight Extensions, IC Tray REF: 888810200
Covidien LP · 2023-08-23
Sterile Procedural Trays, labeled as the following: a. BREAST AUGMENTATION b. BREAST REDUCTION CDS c. CUH ABDOMINO
MEDLINE INDUSTRIES, LP - Northfield · 2023-02-15
VTC Regular Kit Nephrostomy Catheter System Kit, Material Number REF M001245300; to provide external drainage of the uri
Boston Scientific Corporation · 2024-09-25
EnChroma Safety Glasses with Rx Indoor Universal Lenses: Martinez Cx1 Indoor Rx, SKU: Cx.PC.IN.MTZ.BK.Rx; Summit Indoo
Enchroma Inc · 2024-11-27
Otological Ventilation Tube - T-Tube 9mm - Silicone; Product Code: NZ3309;
Adept Medical Ltd · 2025-01-29