RecallHawk
Class I Recall

Baxter EXACTAMIX Inlet, Vented, Micro-Volume Inlet, REF H938175, for use in a pharmaceutical compounding device

Baxter Healthcare Corporation

Summary

The FDA issued a Class I for Baxter EXACTAMIX Inlet, Vented, Micro-Volume Inlet, REF H938175, for use in a ph by Baxter Healthcare Corporation. Reason: Baxter Healthcare Corporation has received increased customer reports of particulate matter in the Automated Compounding Device Inlets (disposable inl.

Details

Source

Device Recall

External ID

Z-3084-2024

Action Date

2024-09-25

Status

Ongoing

Category

device

Product Description

Baxter EXACTAMIX Inlet, Vented, Micro-Volume Inlet, REF H938175, for use in a pharmaceutical compounding device

Lot/Code Info: UDI/DI 00085412475806, Lot Numbers: 803808 and lower

Quantity Affected: 5,828,540 units

Reason for Recall

Baxter Healthcare Corporation has received increased customer reports of particulate matter in the Automated Compounding Device Inlets (disposable inlet), Product Codes H938173, H938174, H938175, H938176, used with the ExactaMix and ExactaMix Pro compounders. Particulate matter has been observed within the inlet primary packaging inlet components, including within the sterile fluid path tubing, before use.

Distribution

Worldwide distribution.

Type: Voluntary: Firm initiated

Recall Initiated: 2024-08-20

Company

Context & Analysis

This is a Class I recall. Class I recalls are the most serious — there is a reasonable probability that use of or exposure to the product will cause serious health consequences or death.

This was one of 257 device recalls issued in the same week, part of 403 device-related FDA actions this month.

Baxter Healthcare Corporation has 310 FDA actions in our database, including 286 recalls and 24 clearances.

Frequently Asked Questions

What is a Class I recall?

A Class I recall is the most serious type. The FDA has determined there is a reasonable probability that the product will cause serious adverse health consequences or death. Consumers should stop using the product immediately.

What should I do if I have this product?

Stop using the product. Check the product against the specific lot numbers or identifiers listed in the recall. Contact the manufacturer (Baxter Healthcare Corporation) or your retailer for a refund or replacement. If you experienced adverse effects, report them to the FDA via MedWatch.

How often does Baxter Healthcare Corporation have FDA actions?

Baxter Healthcare Corporation has 310 FDA actions in our database, including 286 recalls and 24 device clearances.

Where can I find the official FDA record?

The official record is available on FDA.gov. Search for "Z-3084-2024" on FDA.gov for the full record. You can also use the openFDA API to query this record programmatically.

Want to know the moment something like this happens?

Get alerts for device

Related Actions