Baxter EXACTAMIX Inlet, Vented, High Volume Inlet, REF H938174, for use in a pharmaceutical compounding device
Summary
The FDA issued a Class I for Baxter EXACTAMIX Inlet, Vented, High Volume Inlet, REF H938174, for use in a pha by Baxter Healthcare Corporation. Reason: Baxter Healthcare Corporation has received increased customer reports of particulate matter in the Automated Compounding Device Inlets (disposable inl.
Details
Source
Device Recall
External ID
Z-3083-2024
Action Date
2024-09-25
Status
Ongoing
Category
device
Product Description
Baxter EXACTAMIX Inlet, Vented, High Volume Inlet, REF H938174, for use in a pharmaceutical compounding device
Lot/Code Info: UDI/DI 00085412475790, Lot Numbers: 803799 and lower
Quantity Affected: 1,548,782 units
Reason for Recall
Baxter Healthcare Corporation has received increased customer reports of particulate matter in the Automated Compounding Device Inlets (disposable inlet), Product Codes H938173, H938174, H938175, H938176, used with the ExactaMix and ExactaMix Pro compounders. Particulate matter has been observed within the inlet primary packaging inlet components, including within the sterile fluid path tubing, before use.
Distribution
Worldwide distribution.
Type: Voluntary: Firm initiated
Recall Initiated: 2024-08-20
Company
Deerfield, IL
Context & Analysis
This is a Class I recall. Class I recalls are the most serious — there is a reasonable probability that use of or exposure to the product will cause serious health consequences or death.
This was one of 257 device recalls issued in the same week, part of 403 device-related FDA actions this month.
Baxter Healthcare Corporation has 310 FDA actions in our database, including 286 recalls and 24 clearances.
Frequently Asked Questions
What is a Class I recall?
A Class I recall is the most serious type. The FDA has determined there is a reasonable probability that the product will cause serious adverse health consequences or death. Consumers should stop using the product immediately.
What should I do if I have this product?
Stop using the product. Check the product against the specific lot numbers or identifiers listed in the recall. Contact the manufacturer (Baxter Healthcare Corporation) or your retailer for a refund or replacement. If you experienced adverse effects, report them to the FDA via MedWatch.
How often does Baxter Healthcare Corporation have FDA actions?
Baxter Healthcare Corporation has 310 FDA actions in our database, including 286 recalls and 24 device clearances.
Where can I find the official FDA record?
The official record is available on FDA.gov. Search for "Z-3083-2024" on FDA.gov for the full record. You can also use the openFDA API to query this record programmatically.
Want to know the moment something like this happens?
Get alerts for deviceRelated Actions
MAHURKAR 12 Fr High Pressure Triple Lumen Acute Dialysis Catheter, 24 cm, Straight Extensions, IC Tray REF: 888810200
Covidien LP · 2023-08-23
Sterile Procedural Trays, labeled as the following: a. BREAST AUGMENTATION b. BREAST REDUCTION CDS c. CUH ABDOMINO
MEDLINE INDUSTRIES, LP - Northfield · 2023-02-15
VTC Regular Kit Nephrostomy Catheter System Kit, Material Number REF M001245300; to provide external drainage of the uri
Boston Scientific Corporation · 2024-09-25
EnChroma Safety Glasses with Rx Indoor Universal Lenses: Martinez Cx1 Indoor Rx, SKU: Cx.PC.IN.MTZ.BK.Rx; Summit Indoo
Enchroma Inc · 2024-11-27
Otological Ventilation Tube - T-Tube 9mm - Silicone; Product Code: NZ3309;
Adept Medical Ltd · 2025-01-29