RecallHawk
Class II Recall

Medline medical convenience kits, labeled as: 1) LAMINECTOMY PACK, Medline Kit SKU DYNJT6548; 2) ORTHO SPINE PACK,

Medline Industries, LP

Summary

The FDA issued a Class II for Medline medical convenience kits, labeled as: 1) LAMINECTOMY PACK, Medline Ki by Medline Industries, LP. Reason: Specific item(s) and lot(s) of Medline convenience kits containing Pilling Wecksorb Neurosponges and Neurosponge Strips are being recalled due to pote.

Details

Source

Device Recall

External ID

Z-2997-2026

Action Date

2026-09-02

Status

Ongoing

Category

device

Product Description

Medline medical convenience kits, labeled as: 1) LAMINECTOMY PACK, Medline Kit SKU DYNJT6548; 2) ORTHO SPINE PACK, Medline Kit SKU DYNJ44504G; 3) NEURO SPINE PACK, Medline Kit SKU DYNJ45419F; 4) CRANIOTOMY PACK, Medline Kit SKU DYNJ46024F; 5) SPINE PACK, Medline Kit SKU DYNJ54479C; 6) CRANIOTOMY PACK, Medline Kit SKU DYNJ58006B; 7) ORTHO-LAMINECTOMY PACK, Medline Kit SKU DYNJ58014A; 8) SPINE PACK, Medline Kit SKU DYNJ83585; 9) ORTHO SPINE PACK, Medline Kit SKU DYNJ88056; 10) CRANI MT CARMEL, Medline Kit SKU DYNJ900467I; 11) GP-NEURO SPINE PACK-LF, Medline Kit SKU PHS396933009C.

Lot/Code Info: Medline Kit SKU DYNJT6548, UDI/DI 10198459558764 (each) 40198459558765 (case), Lot Number 25JMF561; Medline Kit SKU DYNJ44504G, UDI/DI 10198459241956 (each) 40198459241957 (case), Lot Number 25EDA305; Medline Kit SKU DYNJ44504G, UDI/DI 10198459241956 (each) 40198459241957 (case), Lot Number 25EDB254; Medline Kit SKU DYNJ45419F, UDI/DI 10195327625689 (each) 40195327625680 (case), Lot Number 24JMF418; Medline Kit SKU DYNJ45419F, UDI/DI 10195327625689 (each) 40195327625680 (case), Lot Number 24JMJ129; Medline Kit SKU DYNJ46024F, UDI/DI 10195327237011 (each) 40195327237012 (case), Lot Number 23HBJ773; Medline Kit SKU DYNJ54479C, UDI/DI 10193489460254 (each) 40193489460255 (case), Lot Number 23HBA720; Medline Kit SKU DYNJ58006B, UDI/DI 10193489867381 (each) 40193489867382 (case), Lot Number 23BMC930; Medline Kit SKU DYNJ58006B, UDI/DI 10193489867381 (each) 40193489867382 (case), Lot Number 23FMD505; Medline Kit SKU DYNJ58006B, UDI/DI 10193489867381 (each) 40193489867382 (case), Lot Number 23JMH313; Medline Kit SKU DYNJ58006B, UDI/DI 10193489867381 (each) 40193489867382 (case), Lot Number 23LME066; Medline Kit SKU DYNJ58006B, UDI/DI 10193489867381 (each) 40193489867382 (case), Lot Number 22CMG565; Medline Kit SKU DYNJ58014A, UDI/DI 10193489864007 (each) 40193489864008 (case), Lot Number 24BME326; Medline Kit SKU DYNJ58014A, UDI/DI 10193489864007 (each) 40193489864008 (case), Lot Number 24EMF925; Medline Kit SKU DYNJ58014A, UDI/DI 10193489864007 (each) 40193489864008 (case), Lot Number 23AMJ184; Medline Kit SKU DYNJ58014A, UDI/DI 10193489864007 (each) 40193489864008 (case), Lot Number 23EMD739; Medline Kit SKU DYNJ58014A, UDI/DI 10193489864007 (each) 40193489864008 (case), Lot Number 23IME138; Medline Kit SKU DYNJ58014A, UDI/DI 10193489864007 (each) 40193489864008 (case), Lot Number 23KMB648; Medline Kit SKU DYNJ58014A, UDI/DI 10193489864007 (each) 40193489864008 (case), Lot Number 23LMB107; Medline Kit SKU DYNJ58014A, UDI/DI 10193489864007 (each) 40193489864008 (case), Lot Number 23LMH426; Medline Kit SKU DYNJ58014A, UDI/DI 10193489864007 (each) 40193489864008 (case), Lot Number 22FMF374; Medline Kit SKU DYNJ58014A, UDI/DI 10193489864007 (each) 40193489864008 (case), Lot Number 22KMH219; Medline Kit SKU DYNJ83585, UDI/DI 10195327283179 (each) 40195327283170 (case), Lot Number 23HMC695; Medline Kit SKU DYNJ88056, UDI/DI 10198459092725 (each) 40198459092726 (case), Lot Number 25CMD816; Medline Kit SKU DYNJ88056, UDI/DI 10198459092725 (each) 40198459092726 (case), Lot Number 24IMD021; Medline Kit SKU DYNJ88056, UDI/DI 10198459092725 (each) 40198459092726 (case), Lot Number 24IME838; Medline Kit SKU DYNJ900467I, UDI/DI 10195327181338 (each) 40195327181339 (case), Lot Number 23IBJ627; Medline Kit SKU DYNJ900467I, UDI/DI 10195327181338 (each) 40195327181339 (case), Lot Number 23IBL564; Medline Kit SKU PHS396933009C, UDI/DI 10193489753301 (each) 40193489753302 (case), Lot Number 22EDB595

Quantity Affected: 974 kits

Reason for Recall

Specific item(s) and lot(s) of Medline convenience kits containing Pilling Wecksorb Neurosponges and Neurosponge Strips are being recalled due to potential out-of-specification levels of endotoxin of the neurosponges.

Distribution

US Nationwide distribution in the states of AZ, CA, IA, ID, IL, IN, MA, MD, MI, MN, MO, MT, NY, OH, TX, WI, WY.

Type: Voluntary: Firm initiated

Recall Initiated: 2026-07-08

Company

Context & Analysis

This is a Class II recall. Class II recalls indicate that use of or exposure to the product may cause temporary or medically reversible health consequences, or that the probability of serious consequences is remote.

This was one of 168 device recalls issued in the same week, part of 445 device-related FDA actions this month.

Medline Industries, LP has 577 FDA actions in our database, including 553 recalls and 24 clearances.

Frequently Asked Questions

What is a Class II recall?

A Class II recall means the product may cause temporary or medically reversible adverse health consequences, or the probability of serious adverse consequences is remote. Consumers should discontinue use.

What should I do if I have this product?

Stop using the product. Check the product against the specific lot numbers or identifiers listed in the recall. Contact the manufacturer (Medline Industries, LP) or your retailer for a refund or replacement. If you experienced adverse effects, report them to the FDA via MedWatch.

How often does Medline Industries, LP have FDA actions?

Medline Industries, LP has 577 FDA actions in our database, including 553 recalls and 24 device clearances.

Where can I find the official FDA record?

The official record is available on FDA.gov. Search for "Z-2997-2026" on FDA.gov for the full record. You can also use the openFDA API to query this record programmatically.