RecallHawk
Class II Recall

Beckman Coulter DxC 500 AU Clinical Analyzer Modules using software version 1.5 or lower: 1) AU w/out ISE, Part Numbe

Beckman Coulter Inc.

Summary

The FDA issued a Class II for Beckman Coulter DxC 500 AU Clinical Analyzer Modules using software version 1.5 by Beckman Coulter Inc.. Reason: a software defect in which the Parameter Package (PP) import logic overwrites all predefined calculated test rules, including those previously modif.

Details

Source

Device Recall

External ID

Z-2974-2026

Action Date

2026-08-26

Status

Ongoing

Category

device

Product Description

Beckman Coulter DxC 500 AU Clinical Analyzer Modules using software version 1.5 or lower: 1) AU w/out ISE, Part Number C63519; 2) AU w/ISE, Part Number C63520; DxC 500 AU Clinical Chemistry Analyzers

Lot/Code Info: 1) Part Number C63519, UDI/DI 14987666545058 2) Part Number C63520, UDI/DI 14987666545065

Reason for Recall

a software defect in which the Parameter Package (PP) import logic overwrites all predefined calculated test rules, including those previously modified by the user, upon import of a new Parameter Package version. As a result, usercustomized calculated test rules revert silently to their default expressions, potentially generating erroneous patient test results without an explicit error notification. The issue affects instruments running software version 1.5 or lower and only occurs when a new Parameter Package is imported to a system that already has an existing Parameter Package onboard; new installations are not affected.

Distribution

Worldwide distribution - US Nationwide.

Type: Voluntary: Firm initiated

Recall Initiated: 2026-07-15

Company

Context & Analysis

This is a Class II recall. Class II recalls indicate that use of or exposure to the product may cause temporary or medically reversible health consequences, or that the probability of serious consequences is remote.

This was one of 149 device recalls issued in the same week, part of 445 device-related FDA actions this month.

Beckman Coulter Inc. has 154 FDA actions in our database, including 107 recalls and 47 clearances.

Frequently Asked Questions

What is a Class II recall?

A Class II recall means the product may cause temporary or medically reversible adverse health consequences, or the probability of serious adverse consequences is remote. Consumers should discontinue use.

What should I do if I have this product?

Stop using the product. Check the product against the specific lot numbers or identifiers listed in the recall. Contact the manufacturer (Beckman Coulter Inc.) or your retailer for a refund or replacement. If you experienced adverse effects, report them to the FDA via MedWatch.

How often does Beckman Coulter Inc. have FDA actions?

Beckman Coulter Inc. has 154 FDA actions in our database, including 107 recalls and 47 device clearances.

Where can I find the official FDA record?

The official record is available on FDA.gov. Search for "Z-2974-2026" on FDA.gov for the full record. You can also use the openFDA API to query this record programmatically.

Want to know the moment something like this happens?

Get alerts for device

Related Actions