Summary
The FDA issued a Class II for Impella CP Smart Assist Set; Product Code: 1000402; by Abiomed, Inc.. Reason: A potential device interaction between the Shockwave Coronary IVL Catheter and left-sided Impella Pumps during coronary interventions. Currently there.
Details
Source
Device Recall
External ID
Z-2969-2024
Action Date
2024-09-11
Status
Ongoing
Category
device
Product Description
Impella CP Smart Assist Set; Product Code: 1000402;
Lot/Code Info: Product Code: 1000402; UDI-DI: 00813502012873; Serial Numbers: 385946 385942 385944 385945 385950 385941 385949 385947 385948 385943 392855 392853 392854 392852 392851 392848 392849 392879 392878 427388 427387 427386 430896 430875 430874 430873 430872 430871 430870 430869 430868 430867 431652 430895 430894 486297 486298 486299 486300; Batch Numbers: 2023251705 2023251861 2023251862 2023251869 2023251872 2023251874 2023251876 2023251879 2023251881 2023251902 2023283187 2023283192 2023283194 2023283196 2023283200 2023283208 2023283216 2024305462 2024305463 2024305465 2024305466 2024305467 2024308137 2024308138 2024308139 2024308141 2024308143 2024308144 2024308146 2024308148 2024308149 2024308150 2024308159 2024308160 2024308162 2025501277 2025501292 2025501297 2025501301;
Quantity Affected: 39 units
Reason for Recall
A potential device interaction between the Shockwave Coronary IVL Catheter and left-sided Impella Pumps during coronary interventions. Currently there is no Abiomed labeling (IFU etc.) content related to this potential interaction, so Abiomed will update the IFU to include information on this potential device interaction.
Distribution
Worldwide distribution - US Nationwide and the countries of Australia, Austria, Belgium, Brazil, Brunei Darussalam, Canada, Croatia, Czeck Republic, Denmark, Finland, France, Germany, Hong Kong, India, Ireland, Israel, Italy, Japan, Kuwait, Luxembourg, Malaysia, Mexico, Netherlands, Norway, Panama, Poland, Saudi Arabia, Serbia, Singapore, Slovenia, Spain, Sweden, Switzerland, Taiwan, U.A.E., United Kingdom.
Type: Voluntary: Firm initiated
Recall Initiated: 2024-08-05
Company
Danvers, MA
Context & Analysis
This is a Class II recall. Class II recalls indicate that use of or exposure to the product may cause temporary or medically reversible health consequences, or that the probability of serious consequences is remote.
This was one of 363 device recalls issued in the same week, part of 413 device-related FDA actions this month.
Abiomed, Inc. has 66 FDA actions in our database, including 61 recalls and 5 clearances.
Frequently Asked Questions
What is a Class II recall?
A Class II recall means the product may cause temporary or medically reversible adverse health consequences, or the probability of serious adverse consequences is remote. Consumers should discontinue use.
What should I do if I have this product?
Stop using the product. Check the product against the specific lot numbers or identifiers listed in the recall. Contact the manufacturer (Abiomed, Inc.) or your retailer for a refund or replacement. If you experienced adverse effects, report them to the FDA via MedWatch.
How often does Abiomed, Inc. have FDA actions?
Abiomed, Inc. has 66 FDA actions in our database, including 61 recalls and 5 device clearances.
Where can I find the official FDA record?
The official record is available on FDA.gov. Search for "Z-2969-2024" on FDA.gov for the full record. You can also use the openFDA API to query this record programmatically.
Want to know the moment something like this happens?
Get alerts for deviceRelated Actions
MAHURKAR 12 Fr High Pressure Triple Lumen Acute Dialysis Catheter, 24 cm, Straight Extensions, IC Tray REF: 888810200
Covidien LP · 2023-08-23
Sterile Procedural Trays, labeled as the following: a. BREAST AUGMENTATION b. BREAST REDUCTION CDS c. CUH ABDOMINO
MEDLINE INDUSTRIES, LP - Northfield · 2023-02-15
VTC Regular Kit Nephrostomy Catheter System Kit, Material Number REF M001245300; to provide external drainage of the uri
Boston Scientific Corporation · 2024-09-25
EnChroma Safety Glasses with Rx Indoor Universal Lenses: Martinez Cx1 Indoor Rx, SKU: Cx.PC.IN.MTZ.BK.Rx; Summit Indoo
Enchroma Inc · 2024-11-27
Otological Ventilation Tube - T-Tube 9mm - Silicone; Product Code: NZ3309;
Adept Medical Ltd · 2025-01-29