RecallHawk
Class II Recall

ERBECRYO 2 Cryosurgical Unit Model/Catalog Number: 10402-000

Erbe USA Inc

Summary

The FDA issued a Class II for ERBECRYO 2 Cryosurgical Unit Model/Catalog Number: 10402-000 by Erbe USA Inc. Reason: A strong pull on the high-pressure transfer hose of cryosurgical unit may lead to pneumatic short circuit between he gas inlet and the connector resu.

Details

Source

Device Recall

External ID

Z-2938-2026

Action Date

2026-08-19

Status

Ongoing

Category

device

Product Description

ERBECRYO 2 Cryosurgical Unit Model/Catalog Number: 10402-000

Lot/Code Info: UDI-DI 04050147013797/ Serial Numbers: All serial numbers up to 11608268 Software Version: 1.0.3

Quantity Affected: 11226 units

Reason for Recall

A strong pull on the high-pressure transfer hose of cryosurgical unit may lead to pneumatic short circuit between he gas inlet and the connector resulting in CO2 leaking into the return path, leading to excessive pressure causing the cryoprobe to rupture.

Distribution

US Nationwide distribution including in the states and territories of Alabama, Alaska, Arizona, Arkansas, California, Colorado, Connecticut, Delaware, District of Columbia, Florida, Georgia, Hawaii, Idaho, Illinois, Indiana, Iowa, Kansas, Kentucky, Louisiana, Maine, Maryland, Massachusetts, Michigan, Minnesota, Mississippi, Missouri, Montana, Nebraska, Nevada, New Hampshire, New Jersey, New Mexico, New York, North Carolina, North Dakota, Ohio, Oklahoma, Oregon, Pennsylvania, Puerto Rico, Rhode Island, South Carolina, South Dakota, Tennessee, Texas, Utah, Vermont, Virginia, Washington, West Virginia, Wisconsin, Wyoming.

Type: Voluntary: Firm initiated

Recall Initiated: 2026-06-26

Company

Erbe USA Inc

Marietta, GA

Context & Analysis

This is a Class II recall. Class II recalls indicate that use of or exposure to the product may cause temporary or medically reversible health consequences, or that the probability of serious consequences is remote.

This was one of 160 device recalls issued in the same week, part of 376 device-related FDA actions this month.

Erbe USA Inc has 5 FDA actions in our database, including 4 recalls and 1 clearance.

Frequently Asked Questions

What is a Class II recall?

A Class II recall means the product may cause temporary or medically reversible adverse health consequences, or the probability of serious adverse consequences is remote. Consumers should discontinue use.

What should I do if I have this product?

Stop using the product. Check the product against the specific lot numbers or identifiers listed in the recall. Contact the manufacturer (Erbe USA Inc) or your retailer for a refund or replacement. If you experienced adverse effects, report them to the FDA via MedWatch.

How often does Erbe USA Inc have FDA actions?

Erbe USA Inc has 5 FDA actions in our database, including 4 recalls and 1 device clearance.

Where can I find the official FDA record?

The official record is available on FDA.gov. Search for "Z-2938-2026" on FDA.gov for the full record. You can also use the openFDA API to query this record programmatically.

Want to know the moment something like this happens?

Get alerts for device

Related Actions