RecallHawk
Class II Recall

Neuroguard IEP 3 in 1 Carotid Stent and Post-Dilation Balloon System with Integrated Embolic Protection (Neuroguard IEP

Contego Medical, Inc

Summary

The FDA issued a Class II for Neuroguard IEP 3 in 1 Carotid Stent and Post-Dilation Balloon System with Integr by Contego Medical, Inc. Reason: Carotid stent and post-dilation balloon system with integrated embolic protection system has an integrated filter, that due to a wire-to-filter-bond i.

Details

Source

Device Recall

External ID

Z-2911-2026

Action Date

2026-08-12

Status

Ongoing

Category

device

Product Description

Neuroguard IEP 3 in 1 Carotid Stent and Post-Dilation Balloon System with Integrated Embolic Protection (Neuroguard IEP System), REF: NG-0630-140-2, NG-0640-140-2, NG-0730-140-2, NG-0740-140-2, NG-NV-6-30, NG-NV-6-40, NG-NV-7-30, NG-NV-7-40

Lot/Code Info: REF/UDI-DI/Lot: NG-0630-140-2/00851616007583/Z2460578D,Z2460740B, Z2460872C, Z2460909F, Z2460948D, Z2460957Q; NG-0640-140-2/00851616007590/Z2460578E, Z2460647S, Z2460706A, Z2460746C, Z2460774A, Z2460865C, Z2460872B, Z2460901B, Z2460909D, Z2460909K, Z2460945H, Z2460948I, Z2460963N; NG-0730-140-2/00851616007606/Z2460578F, Z2460662A, Z2460740A , Z2460774F, Z2460850A, Z2460900B, Z2460945F, Z2460957R; NG-0740-140-2/00851616007613/Z2460578G, Z2460635C, Z2460656A, Z2460707D, Z2460733A, Z2460766A, Z2460774B, Z2460780A , Z2460865A, Z2460871A, Z2460872D, Z2460900D, Z2460909B, Z2460909I, Z2460948F, Z2460963T; NG-NV-6-30/00851616007644/Z2460776A, Z2460893A, Z2460909G, Z2460948E; NG-NV-6-40/00851616007651/Z2460776B, Z2460865D, Z2460865E, Z2460901C, Z2460909E, Z2460909J, Z2460957P, Z2460963K, Z2460963P; NG-NV-7-30/00851616007637/Z2460776C, Z2460789H, Z2460850B, Z2460900C, Z2460916A, Z2460948H, Z2460998A; NG-NV-7-40/00851616007620/Z2460776D, Z2460878A, Z2460872A, Z2460872E, Z2460901A, Z2460909C, Z2460909H, Z2460945G, Z2460948G, Z2460966B, Z2460963J;

Quantity Affected: 5069

Reason for Recall

Carotid stent and post-dilation balloon system with integrated embolic protection system has an integrated filter, that due to a wire-to-filter-bond issue, may fail to open following sheath retraction; and if this condition were to occur during a procedure, the result may be loss of intended integrated embolic protection.

Distribution

US Nationwide distribution in the states of IL, FL, NC, WI, KY, AL, VA, DE, NJ, PA, IN, MI, CA.

Type: Voluntary: Firm initiated

Recall Initiated: 2026-06-29

Company

Context & Analysis

This is a Class II recall. Class II recalls indicate that use of or exposure to the product may cause temporary or medically reversible health consequences, or that the probability of serious consequences is remote.

This was one of 156 device recalls issued in the same week, part of 400 device-related FDA actions this month.

Contego Medical, Inc has 6 FDA actions in our database, including 1 recall and 5 clearances.

Frequently Asked Questions

What is a Class II recall?

A Class II recall means the product may cause temporary or medically reversible adverse health consequences, or the probability of serious adverse consequences is remote. Consumers should discontinue use.

What should I do if I have this product?

Stop using the product. Check the product against the specific lot numbers or identifiers listed in the recall. Contact the manufacturer (Contego Medical, Inc) or your retailer for a refund or replacement. If you experienced adverse effects, report them to the FDA via MedWatch.

How often does Contego Medical, Inc have FDA actions?

Contego Medical, Inc has 6 FDA actions in our database, including 1 recall and 5 device clearances.

Where can I find the official FDA record?

The official record is available on FDA.gov. Search for "Z-2911-2026" on FDA.gov for the full record. You can also use the openFDA API to query this record programmatically.

Want to know the moment something like this happens?

Get alerts for device

Related Actions