RecallHawk
Class II Recall

GEM Premier 5000 with IQM2 Model/Catalog Number: 00024019255 and GEM Premier 7000 with iQM3 Model/Catalog Number: 0000

Instrumentation Laboratory

Summary

The FDA issued a Class II for GEM Premier 5000 with IQM2 Model/Catalog Number: 00024019255 and GEM Premier 7 by Instrumentation Laboratory. Reason: Potential for Bar Code Reader (BCR) wands, used with the GEM Premier 5000 and GEM Premier 7000 systems, to assign patient results to an incorrect pati.

Details

Source

Device Recall

External ID

Z-2821-2026

Action Date

2026-08-05

Status

Ongoing

Category

device

Product Description

GEM Premier 5000 with IQM2 Model/Catalog Number: 00024019255 and GEM Premier 7000 with iQM3 Model/Catalog Number: 00000015279 Portable blood analyzer system. Affected Accessory: Bar Code Reader (BCR) Wand Part No. 00024015859

Lot/Code Info: GEM Premier 5000 Part number: 00024019255, UDI: 08426950870227 GEM Premier 7000 Part number: 00000015279, UDI: 0195226000924 Affected Accessory: Bar Code Reader (BCR) Wand that received a configuration update on or after 8 April 2026. Part No. 00024015859; UDI No. 08426950869634

Quantity Affected: 122

Reason for Recall

Potential for Bar Code Reader (BCR) wands, used with the GEM Premier 5000 and GEM Premier 7000 systems, to assign patient results to an incorrect patient ID.

Distribution

US Nationwide distribution.

Type: Voluntary: Firm initiated

Recall Initiated: 2026-05-08

Company

Context & Analysis

This is a Class II recall. Class II recalls indicate that use of or exposure to the product may cause temporary or medically reversible health consequences, or that the probability of serious consequences is remote.

This was one of 136 device recalls issued in the same week, part of 341 device-related FDA actions this month.

Instrumentation Laboratory has 45 FDA actions in our database, including 44 recalls and 1 clearance.

Frequently Asked Questions

What is a Class II recall?

A Class II recall means the product may cause temporary or medically reversible adverse health consequences, or the probability of serious adverse consequences is remote. Consumers should discontinue use.

What should I do if I have this product?

Stop using the product. Check the product against the specific lot numbers or identifiers listed in the recall. Contact the manufacturer (Instrumentation Laboratory) or your retailer for a refund or replacement. If you experienced adverse effects, report them to the FDA via MedWatch.

How often does Instrumentation Laboratory have FDA actions?

Instrumentation Laboratory has 45 FDA actions in our database, including 44 recalls and 1 device clearance.

Where can I find the official FDA record?

The official record is available on FDA.gov. Search for "Z-2821-2026" on FDA.gov for the full record. You can also use the openFDA API to query this record programmatically.

Want to know the moment something like this happens?

Get alerts for device

Related Actions