RecallHawk
Class II Recall

Permobil M3 Corpus Power Wheelchair with Heavy Duty Shock REF:120031, Model version 3

Permobil Inc

Summary

The FDA issued a Class II for Permobil M3 Corpus Power Wheelchair with Heavy Duty Shock REF:120031, Model ver by Permobil Inc. Reason: This recall has been reported retroactively. The power wheelchair firmware lacks a speed restriction to maintain stability when the seat lift exceeds .

Details

Source

Device Recall

External ID

Z-2795-2026

Action Date

2026-08-05

Status

Ongoing

Category

device

Product Description

Permobil M3 Corpus Power Wheelchair with Heavy Duty Shock REF:120031, Model version 3

Lot/Code Info: All Units operating with firmware version 1.26.2/UDI:(01)17330818345674

Quantity Affected: 140 units

Reason for Recall

This recall has been reported retroactively. The power wheelchair firmware lacks a speed restriction to maintain stability when the seat lift exceeds 50 mm, which may cause the user to fall from the wheelchair or the wheelchair to tip over,

Distribution

Worldwide - US Nationwide distribution in the states of AL, AR, AZ, CA, CO, CT, FL, GA, HI, IA, ID, IL, IN, KY, ME, MI, MN, MO, MS, MT, NC, ND, NJ, NV, NY, OH, OK, OR, PA, SC, TN, TX, UT, VA, VT, WA, WI, WV, WY and the country of CAN.

Type: Voluntary: Firm initiated

Recall Initiated: 2024-02-29

Company

Permobil Inc

Lebanon, TN

Context & Analysis

This is a Class II recall. Class II recalls indicate that use of or exposure to the product may cause temporary or medically reversible health consequences, or that the probability of serious consequences is remote.

This was one of 136 device recalls issued in the same week, part of 341 device-related FDA actions this month.

Permobil Inc has 2 FDA actions in our database, including 2 recalls.

Frequently Asked Questions

What is a Class II recall?

A Class II recall means the product may cause temporary or medically reversible adverse health consequences, or the probability of serious adverse consequences is remote. Consumers should discontinue use.

What should I do if I have this product?

Stop using the product. Check the product against the specific lot numbers or identifiers listed in the recall. Contact the manufacturer (Permobil Inc) or your retailer for a refund or replacement. If you experienced adverse effects, report them to the FDA via MedWatch.

How often does Permobil Inc have FDA actions?

Permobil Inc has 2 FDA actions in our database, including 2 recalls and 0 device clearances.

Where can I find the official FDA record?

The official record is available on FDA.gov. Search for "Z-2795-2026" on FDA.gov for the full record. You can also use the openFDA API to query this record programmatically.

Want to know the moment something like this happens?

Get alerts for device

Related Actions