RecallHawk
Class II Recall

Brand Name: Fletcher CT/MR Shielded Applicator Set Product Name: Intracavitary Remote Afterloading Applicator Product

Elekta, Inc.

Summary

The FDA issued a Class II for Brand Name: Fletcher CT/MR Shielded Applicator Set Product Name: Intracavitary by Elekta, Inc.. Reason: Due to safety concerns regarding the detachment of the tip of the intrauterine tube (IU tube) that may remain within a patient..

Details

Source

Device Recall

External ID

Z-2710-2026

Action Date

2026-07-15

Status

Ongoing

Category

device

Product Description

Brand Name: Fletcher CT/MR Shielded Applicator Set Product Name: Intracavitary Remote Afterloading Applicator Product Description: The Fletcher CT/MR Shielded Applicator Set is intended for gynecological brachytherapy treatment of the cervix and endometrium.

Lot/Code Info: Lot Code: 08717213024205

Quantity Affected: 1

Reason for Recall

Due to safety concerns regarding the detachment of the tip of the intrauterine tube (IU tube) that may remain within a patient.

Distribution

Worldwide - U.S. Nationwide distribution including in the states of AR, AZ, CA, CO, CT, DC, FL, GA, IA, IL, IN, KS, KY, LA, MA, MD, MI, MO, MS, NC, ND, NE, NJ, NV, NY, OH, OK, OR, PA, SC, SD, TX, UT, VA, WI, and WV. The countries of Argentina, Australia, Austria, Bahrain, Bangladesh, Belgium, Bolivia, Botswana, Brazil, Burkina Faso, Cambodia, Canada, Chile, China, Colombia, Croatia, Cyprus, Czech Republic, Denmark, Ecuador, Egypt, Finland, France, Georgia, Germany, Greece, Guatemala, Honduras, Hong Kong, Hungary, India, Indonesia, Iran (Islam Republic of), Iraq, Ireland, Italy, Japan, Kazakhstan, Kenya, Korea (Republic of), Kosovo, Malaysia, Mexico, Morocco, Netherlands, New Zealand, Norway, Pakistan, Panama, Peru, Philippines, Poland, Portugal, Romania, Russian Federation, Rwanda, Saudi Arabia, Singapore, Slovakia, South Africa, Spain, Swaziland, Sweden, Switzerland, Taiwan, Thailand, Togo, Turkey, Ukraine, United Arab Emirates, United Kingdom, Venezuela, Viet Nam, and Zambia.

Type: Voluntary: Firm initiated

Recall Initiated: 2026-04-03

Company

Elekta, Inc.

Atlanta, GA

Context & Analysis

This is a Class II recall. Class II recalls indicate that use of or exposure to the product may cause temporary or medically reversible health consequences, or that the probability of serious consequences is remote.

This was one of 143 device recalls issued in the same week, part of 452 device-related FDA actions this month.

Elekta, Inc. has 27 FDA actions in our database, including 27 recalls.

Frequently Asked Questions

What is a Class II recall?

A Class II recall means the product may cause temporary or medically reversible adverse health consequences, or the probability of serious adverse consequences is remote. Consumers should discontinue use.

What should I do if I have this product?

Stop using the product. Check the product against the specific lot numbers or identifiers listed in the recall. Contact the manufacturer (Elekta, Inc.) or your retailer for a refund or replacement. If you experienced adverse effects, report them to the FDA via MedWatch.

How often does Elekta, Inc. have FDA actions?

Elekta, Inc. has 27 FDA actions in our database, including 27 recalls and 0 device clearances.

Where can I find the official FDA record?

The official record is available on FDA.gov. Search for "Z-2710-2026" on FDA.gov for the full record. You can also use the openFDA API to query this record programmatically.

Want to know the moment something like this happens?

Get alerts for device

Related Actions