OTP5000 THORA/PARA TRAY, 5FR, NON-VALVED UDI-DI Code: 0110885403284748 PIG1260T SAFE-T-CENTESIS TRAY 6FR X 16CM UDI
Summary
The FDA issued a Class II for OTP5000 THORA/PARA TRAY, 5FR, NON-VALVED UDI-DI Code: 0110885403284748 PIG12 by Bard Access Systems, Inc.. Reason: Due to lidocaine ampoules being manufactured under deficient manufacturing practices by manufacturer, lidocaine ampoules are currently being recalled .
Details
Source
Device Recall
External ID
Z-2655-2026
Action Date
2026-07-08
Status
Ongoing
Category
device
Product Description
OTP5000 THORA/PARA TRAY, 5FR, NON-VALVED UDI-DI Code: 0110885403284748 PIG1260T SAFE-T-CENTESIS TRAY 6FR X 16CM UDI-DI Code: 0110885403108365 TPT1000 THORACENTESIS/PARACENTESIS TRAY UDI-DI Code: 0110885403108488 The Acute Drainage devices are indicated for use in removing fluid from either the pleural or peritoneal spaces during an acute drainage / aspiration procedure.
Lot/Code Info: OTP5000 THORA/PARA TRAY, 5FR, NON-VALVED UDI-DI Code: 0110885403284748 Lot Number: REKY3402 PIG1260T SAFE-T-CENTESIS TRAY 6FR X 16CM UDI-DI Code: 0110885403108365 Lot Number: REKY3397 TPT1000 THORACENTESIS/PARACENTESIS TRAY UDI-DI Code: 0110885403108488 Lot Number: REKY3386
Quantity Affected: 9,500
Reason for Recall
Due to lidocaine ampoules being manufactured under deficient manufacturing practices by manufacturer, lidocaine ampoules are currently being recalled by supplier.
Distribution
Worldwide U.S. Nationwide distribution including in the states of AK, AL, AR, AZ, CA, CO, CT, DC, DE, FL, GA, HI, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, MT, NC, ND, NE, NH, NJ, NM, NV, NY, OH, OK, OR, PA, PR, RI, SC, SD, TN, TX, UT, VA, WA, WI, WV, and WY. The country of Belgium.
Type: Voluntary: Firm initiated
Recall Initiated: 2026-05-21
Company
Salt Lake City, UT
Context & Analysis
This is a Class II recall. Class II recalls indicate that use of or exposure to the product may cause temporary or medically reversible health consequences, or that the probability of serious consequences is remote.
This was one of 186 device recalls issued in the same week, part of 534 device-related FDA actions this month.
Bard Access Systems, Inc. has 21 FDA actions in our database, including 16 recalls and 5 clearances.
Frequently Asked Questions
What is a Class II recall?
A Class II recall means the product may cause temporary or medically reversible adverse health consequences, or the probability of serious adverse consequences is remote. Consumers should discontinue use.
What should I do if I have this product?
Stop using the product. Check the product against the specific lot numbers or identifiers listed in the recall. Contact the manufacturer (Bard Access Systems, Inc.) or your retailer for a refund or replacement. If you experienced adverse effects, report them to the FDA via MedWatch.
How often does Bard Access Systems, Inc. have FDA actions?
Bard Access Systems, Inc. has 21 FDA actions in our database, including 16 recalls and 5 device clearances.
Where can I find the official FDA record?
The official record is available on FDA.gov. Search for "Z-2655-2026" on FDA.gov for the full record. You can also use the openFDA API to query this record programmatically.
Want to know the moment something like this happens?
Get alerts for deviceRelated Actions
MAHURKAR 12 Fr High Pressure Triple Lumen Acute Dialysis Catheter, 24 cm, Straight Extensions, IC Tray REF: 888810200
Covidien LP · 2023-08-23
Sterile Procedural Trays, labeled as the following: a. BREAST AUGMENTATION b. BREAST REDUCTION CDS c. CUH ABDOMINO
MEDLINE INDUSTRIES, LP - Northfield · 2023-02-15
VTC Regular Kit Nephrostomy Catheter System Kit, Material Number REF M001245300; to provide external drainage of the uri
Boston Scientific Corporation · 2024-09-25
EnChroma Safety Glasses with Rx Indoor Universal Lenses: Martinez Cx1 Indoor Rx, SKU: Cx.PC.IN.MTZ.BK.Rx; Summit Indoo
Enchroma Inc · 2024-11-27
Otological Ventilation Tube - T-Tube 9mm - Silicone; Product Code: NZ3309;
Adept Medical Ltd · 2025-01-29