RecallHawk
Class II Recall

Ivenix Infusion System (IIS), Large Volume Pump (LVP) Software: 5.0.0 and 5.0.1 Part Number: LVP-SW-0004. Indicated

Fresenius Kabi USA, LLC

Summary

The FDA issued a Class II for Ivenix Infusion System (IIS), Large Volume Pump (LVP) Software: 5.0.0 and 5.0.1 by Fresenius Kabi USA, LLC. Reason: Start-up time for the flow rate range greater than 50 ml/hr, and less than 200 ml/hr may exceed the specified 6 minutes, depending on the characterist.

Details

Source

Device Recall

External ID

Z-2635-2023

Action Date

2023-10-04

Status

Ongoing

Category

device

Product Description

Ivenix Infusion System (IIS), Large Volume Pump (LVP) Software: 5.0.0 and 5.0.1 Part Number: LVP-SW-0004. Indicated for use in a hospital and in outpatient care environments for the controlled administration of fluids through clinically accepted routes of administration.

Lot/Code Info: (UDI): 00811505030122 Software Version: 5.0.0 and 5.0.1

Quantity Affected: 1344 pumps

Reason for Recall

Start-up time for the flow rate range greater than 50 ml/hr, and less than 200 ml/hr may exceed the specified 6 minutes, depending on the characteristics of air bubble entrapment within the Intermediate Pumping Chamber (IPC) of the administration set

Distribution

US Nationwide distribution in the states of NJ, WI.

Type: Voluntary: Firm initiated

Recall Initiated: 2022-04-22

Company

Fresenius Kabi USA, LLC

North Andover, MA

Context & Analysis

This is a Class II recall. Class II recalls indicate that use of or exposure to the product may cause temporary or medically reversible health consequences, or that the probability of serious consequences is remote.

This was one of 143 device recalls issued in the same week, part of 403 device-related FDA actions this month.

Fresenius Kabi USA, LLC has 46 FDA actions in our database, including 46 recalls.

Frequently Asked Questions

What is a Class II recall?

A Class II recall means the product may cause temporary or medically reversible adverse health consequences, or the probability of serious adverse consequences is remote. Consumers should discontinue use.

What should I do if I have this product?

Stop using the product. Check the product against the specific lot numbers or identifiers listed in the recall. Contact the manufacturer (Fresenius Kabi USA, LLC) or your retailer for a refund or replacement. If you experienced adverse effects, report them to the FDA via MedWatch.

How often does Fresenius Kabi USA, LLC have FDA actions?

Fresenius Kabi USA, LLC has 46 FDA actions in our database, including 46 recalls and 0 device clearances.

Where can I find the official FDA record?

The official record is available on FDA.gov. Search for "Z-2635-2023" on FDA.gov for the full record. You can also use the openFDA API to query this record programmatically.