PDS Plus Antibacterial (polydioxanone) Suture, Product Numbers PDP072H40, PDP259H50, PDP416H50, PDP441H50, PDP442H50, PD
Summary
The FDA issued a Class II for PDS Plus Antibacterial (polydioxanone) Suture, Product Numbers PDP072H40, PDP259 by ETHICON, LLC. Reason: Internal testing on returned product from this lot confirmed that some PDS" II (polydioxanone) Sutures from this lot did not meet tensile strength req.
Details
Source
Device Recall
External ID
Z-2616-2023
Action Date
2023-09-27
Status
Ongoing
Category
device
Product Description
PDS Plus Antibacterial (polydioxanone) Suture, Product Numbers PDP072H40, PDP259H50, PDP416H50, PDP441H50, PDP442H50, PDP489G50, PDP489H40, PDP497G51, PDP497H16, PDP497H40, PDP498H40, PDP502V05, PDP507V05, PDP510V05, PDP515V05, PDP662H50, PDP663H50, PDP684H40, PDP9615H40, PDP9625H40, PDP9626H40, PDP9631H40, PDP9715H15, PDP9715H40, PDP9733H16, PDP9861H40, PDP9865H40, XWPDP421H40, XWPDP423H40, XWPDP442H40, XWPDP443H16
Lot/Code Info: UDI-DI: 10705031122574, 10705031047952, 10705031048232, 10705031048294, 10705031048300, 10705031048393, 10705031123786, 10705031048447, 10705031123854, 10705031123854, 10705031123861, 10705031203969, 10705031203983, 10705031204010, 10705031204034, 10705031048546, 10705031048553, 10705031124059, 10705031124677, 10705031124691, 10705031124707, 10705031124714, 10705031124752, 10705031124752, 10705031124769, 10705031124875, 10705031124899, 10705031123540, 10705031123564, 10705031123588, 10705031123595; Lot Numbers: RLMCKU, RLMEAL, RLMDST, RMMHLT, RJMMAC, SCMASA, RKMQDX, RKMBEB, SAMERU
Quantity Affected: 868,363 total units
Reason for Recall
Internal testing on returned product from this lot confirmed that some PDS" II (polydioxanone) Sutures from this lot did not meet tensile strength requirement.
Distribution
US Nationwide. Global Distribution.
Type: Voluntary: Firm initiated
Recall Initiated: 2023-08-10
Company
Guaynabo, PR
Context & Analysis
This is a Class II recall. Class II recalls indicate that use of or exposure to the product may cause temporary or medically reversible health consequences, or that the probability of serious consequences is remote.
This was one of 195 device recalls issued in the same week, part of 403 device-related FDA actions this month.
ETHICON, LLC has 6 FDA actions in our database, including 6 recalls.
Frequently Asked Questions
What is a Class II recall?
A Class II recall means the product may cause temporary or medically reversible adverse health consequences, or the probability of serious adverse consequences is remote. Consumers should discontinue use.
What should I do if I have this product?
Stop using the product. Check the product against the specific lot numbers or identifiers listed in the recall. Contact the manufacturer (ETHICON, LLC) or your retailer for a refund or replacement. If you experienced adverse effects, report them to the FDA via MedWatch.
How often does ETHICON, LLC have FDA actions?
ETHICON, LLC has 6 FDA actions in our database, including 6 recalls and 0 device clearances.
Where can I find the official FDA record?
The official record is available on FDA.gov. Search for "Z-2616-2023" on FDA.gov for the full record. You can also use the openFDA API to query this record programmatically.
Want to know the moment something like this happens?
Get alerts for deviceRelated Actions
MAHURKAR 12 Fr High Pressure Triple Lumen Acute Dialysis Catheter, 24 cm, Straight Extensions, IC Tray REF: 888810200
Covidien LP · 2023-08-23
Sterile Procedural Trays, labeled as the following: a. BREAST AUGMENTATION b. BREAST REDUCTION CDS c. CUH ABDOMINO
MEDLINE INDUSTRIES, LP - Northfield · 2023-02-15
VTC Regular Kit Nephrostomy Catheter System Kit, Material Number REF M001245300; to provide external drainage of the uri