Reprocessed Webster CS Catheter 10 Electrodes for electrophysiological mapping of cardiac structures. Medline Item Numbe

Medline Industries, LP

Summary

The FDA issued a Class I for Reprocessed Webster CS Catheter 10 Electrodes for electrophysiological mapping o by Medline Industries, LP. Reason: Specific lots of reprocessed electrophysiology and ultrasound catheters may contain small residual particulates. If devices with residual particulates.

Details

Source

Device Recall

External ID

Z-2611-2025

Action Date

2025-10-08

Status

Ongoing

Category

device

Product Description

Reprocessed Webster CS Catheter 10 Electrodes for electrophysiological mapping of cardiac structures. Medline Item Numbers BD710DF282CRH BD710DF282RRH BD710FJ282CRH BD710FJ282RRH

Lot/Code Info: Item Number/UDI-DI BD710DF282CRH 10197344044023; BD710DF282RRH 10197344044030; BD710FJ282CRH 10197344044047; BD710FJ282RRH 10197344044054; LOTS EP241216 EP241218 EP241226 EP241231 EP250110 EP250121 EP250210

Quantity Affected: 7913 eaches

Reason for Recall

Specific lots of reprocessed electrophysiology and ultrasound catheters may contain small residual particulates. If devices with residual particulates on patient-contacting surfaces are used, there is a potential risk of an inflammatory response or systemic infection, and if particulate were to dislodge during use and contact circulating blood and/or the heart, it may produce a granulomatous reaction or thrombus formation, which may lead to cerebral or pulmonary embolism, or deep vein thrombosis.

Distribution

Domestic US distribution nationwide. International distribution pending. No international distribution.

Type: Voluntary: Firm initiated

Recall Initiated: 2025-07-31

Company

Context & Analysis

This is a Class I recall. Class I recalls are the most serious — there is a reasonable probability that use of or exposure to the product will cause serious health consequences or death.

This was one of 168 device recalls issued in the same week, part of 413 device-related FDA actions this month.

Medline Industries, LP has 442 FDA actions in our database, including 415 recalls and 27 clearances.

Frequently Asked Questions

What is a Class I recall?

A Class I recall is the most serious type. The FDA has determined there is a reasonable probability that the product will cause serious adverse health consequences or death. Consumers should stop using the product immediately.

What should I do if I have this product?

Stop using the product. Check the product against the specific lot numbers or identifiers listed in the recall. Contact the manufacturer (Medline Industries, LP) or your retailer for a refund or replacement. If you experienced adverse effects, report them to the FDA via MedWatch.

How often does Medline Industries, LP have FDA actions?

Medline Industries, LP has 442 FDA actions in our database, including 415 recalls and 27 device clearances.

Where can I find the official FDA record?

The official record is available on FDA.gov. Search for "Z-2611-2025" on FDA.gov for the full record. You can also use the openFDA API to query this record programmatically.

Want to know the moment something like this happens?

Get alerts for device

Related Actions