RecallHawk
Class II Recall

Medline Convenience kits containing recalled Swan-Ganz Catheters CARDIAC MAJOR DYNJ901328G HEART A DYNJ903

Medline Industries, LP

Summary

The FDA issued a Class II for Medline Convenience kits containing recalled Swan-Ganz Catheters CARDIAC MAJO by Medline Industries, LP. Reason: Kits contain Swan-Ganz catheters recalled by Becton Dickinson (BD). BD has received complaints of leaks/breakage involving failures of the (blue) prox.

Details

Source

Device Recall

External ID

Z-2607-2026

Action Date

2026-07-08

Status

Ongoing

Category

device

Product Description

Medline Convenience kits containing recalled Swan-Ganz Catheters CARDIAC MAJOR DYNJ901328G HEART A DYNJ903465K OPEN HEART DYNJ909090G SILVER CROSS OPEN HEART DYNJ906102M VPH OPEN HEART DYNJ902664M DYNJ902664N

Lot/Code Info: DYNJ901328G UDI-DI 10198459601538 Lots 26ABD786 26CBD768 26DBJ434 DYNJ903465K UDI-DI 10198459292743 Lots 25GDB257 26ADA464 26BDA587 26DDA021 DYNJ909090G UDI-DI 10198459406027 Lots 25LBH690 25LBP697 DYNJ906102M UDI-DI 10198459406409 Lots 25JBE315 25KBI700 25KBM827 25LBB028 26ABH106 26CBQ206 DYNJ902664M UDI-DI 10198459349126 Lots 25ILA177 25KLA500 DYNJ902664N UDI-DI 10198459675287 Lot 26BLA783

Quantity Affected: 9,272 kits

Reason for Recall

Kits contain Swan-Ganz catheters recalled by Becton Dickinson (BD). BD has received complaints of leaks/breakage involving failures of the (blue) proximal injectate lumen hub on these catheters. This has led to instances of leakage, lumen damage, and breakage, and may lead to infection, medication loss, and/or blood loss.

Distribution

US Nationwide distribution in the states of AZ, CA, FL, IL, MO, NJ, OH, OK, TN, TX, WI.

Type: Voluntary: Firm initiated

Recall Initiated: 2026-05-28

Company

Context & Analysis

This is a Class II recall. Class II recalls indicate that use of or exposure to the product may cause temporary or medically reversible health consequences, or that the probability of serious consequences is remote.

This was one of 186 device recalls issued in the same week, part of 534 device-related FDA actions this month.

Medline Industries, LP has 487 FDA actions in our database, including 459 recalls and 28 clearances.

Frequently Asked Questions

What is a Class II recall?

A Class II recall means the product may cause temporary or medically reversible adverse health consequences, or the probability of serious adverse consequences is remote. Consumers should discontinue use.

What should I do if I have this product?

Stop using the product. Check the product against the specific lot numbers or identifiers listed in the recall. Contact the manufacturer (Medline Industries, LP) or your retailer for a refund or replacement. If you experienced adverse effects, report them to the FDA via MedWatch.

How often does Medline Industries, LP have FDA actions?

Medline Industries, LP has 487 FDA actions in our database, including 459 recalls and 28 device clearances.

Where can I find the official FDA record?

The official record is available on FDA.gov. Search for "Z-2607-2026" on FDA.gov for the full record. You can also use the openFDA API to query this record programmatically.