RecallHawk
Class II Recall

Atomlab 500 Dose Calibrator Software 2.0.00 through Revision 2.0.08, inclusive Model: 086-330.

Biodex Medical Systems, Inc.

Summary

The FDA issued a Class II for Atomlab 500 Dose Calibrator Software 2.0.00 through Revision 2.0.08, inclusive M by Biodex Medical Systems, Inc.. Reason: When deleting a previously entered custom isotope, the software deletes the isotope, but not the associated dial setting. Possible other stored custom.

Details

Source

Device Recall

External ID

Z-2580-2021

Action Date

2021-10-06

Status

Terminated

Category

device

Product Description

Atomlab 500 Dose Calibrator Software 2.0.00 through Revision 2.0.08, inclusive Model: 086-330.

Lot/Code Info: Serial Number range:17081557-21062362 UDI: 00718175003305

Quantity Affected: 829 units

Reason for Recall

When deleting a previously entered custom isotope, the software deletes the isotope, but not the associated dial setting. Possible other stored custom isotopes may then have incorrect dial settings and may lead to an incorrect dose measurement and/or misadministration

Distribution

Nationwide Foreign: Austria Autralia Brazil CANADA Chile Colombia Croatia Cyprus Czech Republic Denmark Dominican Republic Ecuador Egypt France Hong Kong HUNGARY Iceland India Iraq Ireland ISRAEL Isreal Italy Japan JORDAN Kuwait Lebanon Libya Malaysia MALTA Mexico Morocco Netherlands New Zealand Oman Pakistan Palestine Panama Peru Philippines Poland Portugal PR Puerto Rico Qatar Romania Russia Saudi Arabia Singapore South Africa South Korea Spain Sweden Taiwan UAE United Arab Emirates United Kingdom Venezuela

Type: Voluntary: Firm initiated

Recall Initiated: 2021-08-16

Context & Analysis

This is a Class II recall. Class II recalls indicate that use of or exposure to the product may cause temporary or medically reversible health consequences, or that the probability of serious consequences is remote.

This was one of 201 device recalls issued in the same week, part of 403 device-related FDA actions this month.

Biodex Medical Systems, Inc. has 2 FDA actions in our database, including 2 recalls.

Frequently Asked Questions

What is a Class II recall?

A Class II recall means the product may cause temporary or medically reversible adverse health consequences, or the probability of serious adverse consequences is remote. Consumers should discontinue use.

What should I do if I have this product?

Stop using the product. Check the product against the specific lot numbers or identifiers listed in the recall. Contact the manufacturer (Biodex Medical Systems, Inc.) or your retailer for a refund or replacement. If you experienced adverse effects, report them to the FDA via MedWatch.

How often does Biodex Medical Systems, Inc. have FDA actions?

Biodex Medical Systems, Inc. has 2 FDA actions in our database, including 2 recalls and 0 device clearances.

Where can I find the official FDA record?

The official record is available on FDA.gov. Search for "Z-2580-2021" on FDA.gov for the full record. You can also use the openFDA API to query this record programmatically.