Chemistry Products LAC Slides. Catalog Number: 8433880. 1 Sales Unit = contains 5 cartridges of 60 slides, for a total o
Summary
The FDA issued a Class II for Chemistry Products LAC Slides. Catalog Number: 8433880. 1 Sales Unit = contains by Ortho-Clinical Diagnostics, Inc.. Reason: VITROS Chemistry Products LAC Slides on VITROS XT 3400 and VITROS XT 7600 Systems are experiencing an increased occurrence of calibration failure rela.
Details
Source
Device Recall
External ID
Z-2559-2025
Action Date
2025-09-24
Status
Ongoing
Category
device
Product Description
Chemistry Products LAC Slides. Catalog Number: 8433880. 1 Sales Unit = contains 5 cartridges of 60 slides, for a total of 300 tests.
Lot/Code Info: Catalog number 8433880; UDI# 110758750004911; Coating 0130 and above, this includes Lot numbers 3533-0130-2317, 3533-0130-2717, 3533-0130-2718, 3533-0130-2719, 3533-0130-4546, 3533-0130-4605, 3533-0131-4763, 3533-0131-6114, 3533-0131-6115, 3533-0131-6116, 3533-0131-6128; Expiration date range 01-Sep-2026 to 01-Jan-2027
Quantity Affected: 2,908 units
Reason for Recall
VITROS Chemistry Products LAC Slides on VITROS XT 3400 and VITROS XT 7600 Systems are experiencing an increased occurrence of calibration failure related to lots from coating 0130 and above with Condition Code TH4-63J during calibration, causing the delay of lac test results and potentially impact patient management. The issue was identified by QuidelOrtho through an increase in customer complaints compared to the historical baseline with average failure rate of 1% for coating 0130 in 2025 while average of failure rate for other coating was 0.06% in 2025 and average of failure rate for all coatings is 0.03% for 2024.
Distribution
US Nationwide distribution.
Type: Voluntary: Firm initiated
Recall Initiated: 2025-07-29
Company
Rochester, NY
Context & Analysis
This is a Class II recall. Class II recalls indicate that use of or exposure to the product may cause temporary or medically reversible health consequences, or that the probability of serious consequences is remote.
This was one of 133 device recalls issued in the same week, part of 403 device-related FDA actions this month.
Ortho-Clinical Diagnostics, Inc. has 71 FDA actions in our database, including 67 recalls and 4 clearances.
Frequently Asked Questions
What is a Class II recall?
A Class II recall means the product may cause temporary or medically reversible adverse health consequences, or the probability of serious adverse consequences is remote. Consumers should discontinue use.
What should I do if I have this product?
Stop using the product. Check the product against the specific lot numbers or identifiers listed in the recall. Contact the manufacturer (Ortho-Clinical Diagnostics, Inc.) or your retailer for a refund or replacement. If you experienced adverse effects, report them to the FDA via MedWatch.
How often does Ortho-Clinical Diagnostics, Inc. have FDA actions?
Ortho-Clinical Diagnostics, Inc. has 71 FDA actions in our database, including 67 recalls and 4 device clearances.
Where can I find the official FDA record?
The official record is available on FDA.gov. Search for "Z-2559-2025" on FDA.gov for the full record. You can also use the openFDA API to query this record programmatically.
Want to know the moment something like this happens?
Get alerts for deviceRelated Actions
MAHURKAR 12 Fr High Pressure Triple Lumen Acute Dialysis Catheter, 24 cm, Straight Extensions, IC Tray REF: 888810200
Covidien LP · 2023-08-23
Sterile Procedural Trays, labeled as the following: a. BREAST AUGMENTATION b. BREAST REDUCTION CDS c. CUH ABDOMINO
MEDLINE INDUSTRIES, LP - Northfield · 2023-02-15
VTC Regular Kit Nephrostomy Catheter System Kit, Material Number REF M001245300; to provide external drainage of the uri
Boston Scientific Corporation · 2024-09-25
EnChroma Safety Glasses with Rx Indoor Universal Lenses: Martinez Cx1 Indoor Rx, SKU: Cx.PC.IN.MTZ.BK.Rx; Summit Indoo
Enchroma Inc · 2024-11-27
Otological Ventilation Tube - T-Tube 9mm - Silicone; Product Code: NZ3309;