RecallHawk
Class II Recall

Hookup Model/Catalog Number: 2-8-540HAN Product Description: Hookups provide the interface between an Advantage Plus

Steris Corporation

Summary

The FDA issued a Class II for Hookup Model/Catalog Number: 2-8-540HAN Product Description: Hookups provide by Steris Corporation. Reason: IFU for hookup were incorrectly reflected within hookup IFUs and online reference system, Hookup Lookup (HULU). Use of an incorrect hookup connection .

Details

Source

Device Recall

External ID

Z-2537-2024

Action Date

2024-08-14

Status

Ongoing

Category

device

Product Description

Hookup Model/Catalog Number: 2-8-540HAN Product Description: Hookups provide the interface between an Advantage Plus" Automated Endoscope Reprocessor and Advantage Plus" Pass-Thru Automated Endoscope Reprocessor ( AER ) and endoscopes, which allow perfusion of liquids dispensed by the AER to contact and high-level disinfect the channels of flexible endoscopes.

Lot/Code Info: Lot Code: UDI-DI 00677964086885; Lot Numbers 415308, 415665, 422921, 456865, 458947, 461749, 463591, 465821, 467121, 467705, 468813, 480858, 484468, 513285, 517588, 533998, 544724, 585095, 617851

Quantity Affected: 129

Reason for Recall

IFU for hookup were incorrectly reflected within hookup IFUs and online reference system, Hookup Lookup (HULU). Use of an incorrect hookup connection and/or parameter set may prevent an endoscope from being properly disinfected, which may present a risk of patient contamination or infection

Distribution

Worldwide distribution - United States Nationwide and the countries of Canada, Belgium, Czech Republic, Denmark, Germany, Spain, Italy, Slovenia, Slovakia, Australia, Israel, Jordan, Kuwait, Namibia, Saudi Arabia, South Africa, Cote d Ivoire, Afghanistan, Albania, Belarus, Botswana, Kazakhstan, Malaysia, Qatar, Ukraine, United Arab Emirates, Nicaragua, and Nigeria.

Type: Voluntary: Firm initiated

Recall Initiated: 2024-06-13

Company

Context & Analysis

This is a Class II recall. Class II recalls indicate that use of or exposure to the product may cause temporary or medically reversible health consequences, or that the probability of serious consequences is remote.

This was one of 158 device recalls issued in the same week, part of 403 device-related FDA actions this month.

Steris Corporation has 71 FDA actions in our database, including 28 recalls and 43 clearances.

Frequently Asked Questions

What is a Class II recall?

A Class II recall means the product may cause temporary or medically reversible adverse health consequences, or the probability of serious adverse consequences is remote. Consumers should discontinue use.

What should I do if I have this product?

Stop using the product. Check the product against the specific lot numbers or identifiers listed in the recall. Contact the manufacturer (Steris Corporation) or your retailer for a refund or replacement. If you experienced adverse effects, report them to the FDA via MedWatch.

How often does Steris Corporation have FDA actions?

Steris Corporation has 71 FDA actions in our database, including 28 recalls and 43 device clearances.

Where can I find the official FDA record?

The official record is available on FDA.gov. Search for "Z-2537-2024" on FDA.gov for the full record. You can also use the openFDA API to query this record programmatically.

Want to know the moment something like this happens?

Get alerts for device

Related Actions