RecallHawk
Class II Recall

Boston Scientific Encore 26 Inflation Device, Material Numbers (UPN): 1. H74904526011; 2. H74904526052; 3. M00671

Boston Scientific Corporation

Summary

The FDA issued a Class II for Boston Scientific Encore 26 Inflation Device, Material Numbers (UPN): 1. H749 by Boston Scientific Corporation. Reason: The Encore 26 Inflation Device is used with balloon dilation catheters to create and monitor pressure in the balloon and to deflate the balloon. An in.

Details

Source

Device Recall

External ID

Z-2533-2025

Action Date

2025-09-17

Status

Ongoing

Category

device

Product Description

Boston Scientific Encore 26 Inflation Device, Material Numbers (UPN): 1. H74904526011; 2. H74904526052; 3. M0067101140; 4. M001151050; 5. M001151062; 6. M00566670. (Interventional Cardiology, Peripheral Interventions)

Lot/Code Info: 1. H74904526011, GTIN 8714729177029, Lot Numbers: 34915076, 35065978; 2. H74904526052, GTIN 8714729127062, Lot Numbers: 34915078, 34915220; 3. M0067101140, GTIN 8714729755814, Lot Numbers: 34892421; 4. M001151050, GTIN 8714729183624, Lot Numbers: 34966934; 5. M001151062, GTIN 8714729137542, Lot Numbers: 34873498; 6. M00566670, GTIN 8714729755241, Lot Numbers: 35045668.

Quantity Affected: 39739 units

Reason for Recall

The Encore 26 Inflation Device is used with balloon dilation catheters to create and monitor pressure in the balloon and to deflate the balloon. An internal investigation determined that during use, foreign material particles could migrate from affected Encore 26 Inflation Devices into a balloon dilatation catheter.

Distribution

Worldwide distribution.

Type: Voluntary: Firm initiated

Recall Initiated: 2025-08-05

Company

Context & Analysis

This is a Class II recall. Class II recalls indicate that use of or exposure to the product may cause temporary or medically reversible health consequences, or that the probability of serious consequences is remote.

This was one of 168 device recalls issued in the same week, part of 413 device-related FDA actions this month.

Boston Scientific Corporation has 399 FDA actions in our database, including 334 recalls and 65 clearances.

Frequently Asked Questions

What is a Class II recall?

A Class II recall means the product may cause temporary or medically reversible adverse health consequences, or the probability of serious adverse consequences is remote. Consumers should discontinue use.

What should I do if I have this product?

Stop using the product. Check the product against the specific lot numbers or identifiers listed in the recall. Contact the manufacturer (Boston Scientific Corporation) or your retailer for a refund or replacement. If you experienced adverse effects, report them to the FDA via MedWatch.

How often does Boston Scientific Corporation have FDA actions?

Boston Scientific Corporation has 399 FDA actions in our database, including 334 recalls and 65 device clearances.

Where can I find the official FDA record?

The official record is available on FDA.gov. Search for "Z-2533-2025" on FDA.gov for the full record. You can also use the openFDA API to query this record programmatically.