Summary

The FDA issued a Class II for iLet Bionic Pancreas, REF: BB1001 by Beta Bionics, Inc.. Reason: The reason for the recall is: it was found in version 3.3.6 of the iLet ACE Pump firmware and all previous releases that a screen unlock failure could.

Details

Source

Device Recall

External ID

Z-2513-2026

Action Date

2026-06-24

Status

Ongoing

Category

device

Product Description

iLet Bionic Pancreas, REF: BB1001

Lot/Code Info: iLet ACE Pump firmware BOM000072 Rev C (version 3.3.6) and BOM000072 Rev B (version 3.3.4)

Quantity Affected: 4580

Reason for Recall

The reason for the recall is: it was found in version 3.3.6 of the iLet ACE Pump firmware and all previous releases that a screen unlock failure could occur when the user successfully slides the unlock slider on the Lock Screen, no screen transition occurs, and the user remains on the Lock Screen. Because the user cannot exit the Lock Screen, the user is unable to perform a cartridge change or meal announcement. If this occurs and therapy is in progress, therapy will continue normally until the insulin cartridge is depleted. The user will be unable to perform a cartridge change or a meal announcement and will be stuck on the screen until the battery fully drains. This can lead to hyperglycemia.

Distribution

US: WI, HI, OH, ME, CO, NJ, NE, KY, SC, CA, TX, NH, NY, WA, PA, MD, FL, IA, GA, MI, VT, TN, IL, IN, KS, MA, LA, MS, NC, CT, MN, VA, OK, AR, AZ, MO, MT, RI, DC, NM, OR, ID, NV, AK, ND, DE, WV, WY, SD, UT, AL

Type: Voluntary: Firm initiated

Recall Initiated: 2025-09-23

Company

Context & Analysis

This is a Class II recall. Class II recalls indicate that use of or exposure to the product may cause temporary or medically reversible health consequences, or that the probability of serious consequences is remote.

This was one of 254 device recalls issued in the same week, part of 403 device-related FDA actions this month.

Beta Bionics, Inc. has 8 FDA actions in our database, including 3 recalls and 5 clearances.

Frequently Asked Questions

What is a Class II recall?

A Class II recall means the product may cause temporary or medically reversible adverse health consequences, or the probability of serious adverse consequences is remote. Consumers should discontinue use.

What should I do if I have this product?

Stop using the product. Check the product against the specific lot numbers or identifiers listed in the recall. Contact the manufacturer (Beta Bionics, Inc.) or your retailer for a refund or replacement. If you experienced adverse effects, report them to the FDA via MedWatch.

How often does Beta Bionics, Inc. have FDA actions?

Beta Bionics, Inc. has 8 FDA actions in our database, including 3 recalls and 5 device clearances.

Where can I find the official FDA record?

The official record is available on FDA.gov. Search for "Z-2513-2026" on FDA.gov for the full record. You can also use the openFDA API to query this record programmatically.

Want to know the moment something like this happens?

Get alerts for device

Related Actions