RecallHawk
Class II Recall

RefleXion X1 Model RXM1000 UDI Codes: (01)00860003983812(11)230504(21)X11001 (01)00860003983812(11)230518(21)X11002

Reflexion Medical, Inc.

Summary

The FDA issued a Class II for RefleXion X1 Model RXM1000 UDI Codes: (01)00860003983812(11)230504(21)X11001 by Reflexion Medical, Inc.. Reason: Due to incorrect software configuration that potentially allows more than two (2) fractions within a 12-hour period (fractionLimits) and could potenti.

Details

Source

Device Recall

External ID

Z-2486-2025

Action Date

2025-09-10

Status

Ongoing

Category

device

Product Description

RefleXion X1 Model RXM1000 UDI Codes: (01)00860003983812(11)230504(21)X11001 (01)00860003983812(11)230518(21)X11002 (01)00860003983812(11)230505(21)X11003 (01)00860003983812(11)230530(21)X11004 (01)00860003983812(11)230811(21)X11006 (01)00860003983812(11)230410(21)X11007 (01)00860003983812(11)230803(21)X11008 (01)00860003983812(11)240528(21)X11009 (01)00860003983812(11)240820(21)X11010 (01)00860003983812(11)241107(21)X11011 (01)00860003983812(11)241206(21)X11012 The RefleXion X1 is a radiation therapy system that combines the functionality of an emission computed tomography detection system and a linear accelerator. The device is intended for use with approved fludeoxyglucose F18. The emission computed tomography detection system acquires images of positron-emitting fludeoxyglucose F18 for the purpose of guiding the delivery of megavoltage X-rays for oncologic treatment with radiation therapy using an FDA cleared, authorized, or approved linear accelerator.

Lot/Code Info: Model Number: RXM1000 UDI Codes that includes Serial Numbers: (01)00860003983812(11)230504(21)X11001 (01)00860003983812(11)230518(21)X11002 (01)00860003983812(11)230505(21)X11003 (01)00860003983812(11)230530(21)X11004 (01)00860003983812(11)230811(21)X11006 (01)00860003983812(11)230410(21)X11007 (01)00860003983812(11)230803(21)X11008 (01)00860003983812(11)240528(21)X11009 (01)00860003983812(11)240820(21)X11010 (01)00860003983812(11)241107(21)X11011 (01)00860003983812(11)241206(21)X11012

Quantity Affected: 11

Reason for Recall

Due to incorrect software configuration that potentially allows more than two (2) fractions within a 12-hour period (fractionLimits) and could potentially lead to adverse events (toxicity).

Distribution

U.S. Nationwide distribution in the states of CA, CT, LA, NJ, OR, PA, and TX.

Type: Voluntary: Firm initiated

Recall Initiated: 2025-07-22

Company

Context & Analysis

This is a Class II recall. Class II recalls indicate that use of or exposure to the product may cause temporary or medically reversible health consequences, or that the probability of serious consequences is remote.

This was one of 258 device recalls issued in the same week, part of 403 device-related FDA actions this month.

Reflexion Medical, Inc. has 6 FDA actions in our database, including 4 recalls and 2 clearances.

Frequently Asked Questions

What is a Class II recall?

A Class II recall means the product may cause temporary or medically reversible adverse health consequences, or the probability of serious adverse consequences is remote. Consumers should discontinue use.

What should I do if I have this product?

Stop using the product. Check the product against the specific lot numbers or identifiers listed in the recall. Contact the manufacturer (Reflexion Medical, Inc.) or your retailer for a refund or replacement. If you experienced adverse effects, report them to the FDA via MedWatch.

How often does Reflexion Medical, Inc. have FDA actions?

Reflexion Medical, Inc. has 6 FDA actions in our database, including 4 recalls and 2 device clearances.

Where can I find the official FDA record?

The official record is available on FDA.gov. Search for "Z-2486-2025" on FDA.gov for the full record. You can also use the openFDA API to query this record programmatically.