Summary
The FDA issued a Class II for GE HealthCare OEC 9900 Elite C-arm Systems. by GE Medical Systems, LLC. Reason: GE HealthCare has become aware of a non-conformance to 21 CFR 1020.30 (k) in certain OEC 9800 and OEC 9900 Elite C-arm Systems. Specifically, impacted.
Details
Source
Device Recall
External ID
Z-2342-2025
Action Date
2025-09-03
Status
Ongoing
Category
device
Product Description
GE HealthCare OEC 9900 Elite C-arm Systems.
Lot/Code Info: GTIN: 00840682114349
Reason for Recall
GE HealthCare has become aware of a non-conformance to 21 CFR 1020.30 (k) in certain OEC 9800 and OEC 9900 Elite C-arm Systems. Specifically, impacted systems are missing the full layer of lead tape in the collimator covers, which will lead to a small increase in radiation exposure to the patient and/or operator. The missing lead tape is not detectable by the user.
Distribution
US Nationwide distribution.
Type: FDA Mandated
Recall Initiated: 2025-07-02
Company
Waukesha, WI
Context & Analysis
This is a Class II recall. Class II recalls indicate that use of or exposure to the product may cause temporary or medically reversible health consequences, or that the probability of serious consequences is remote.
This was one of 201 device recalls issued in the same week, part of 403 device-related FDA actions this month.
GE Medical Systems, LLC has 138 FDA actions in our database, including 121 recalls and 17 clearances.
Frequently Asked Questions
What is a Class II recall?
A Class II recall means the product may cause temporary or medically reversible adverse health consequences, or the probability of serious adverse consequences is remote. Consumers should discontinue use.
What should I do if I have this product?
Stop using the product. Check the product against the specific lot numbers or identifiers listed in the recall. Contact the manufacturer (GE Medical Systems, LLC) or your retailer for a refund or replacement. If you experienced adverse effects, report them to the FDA via MedWatch.
How often does GE Medical Systems, LLC have FDA actions?
GE Medical Systems, LLC has 138 FDA actions in our database, including 121 recalls and 17 device clearances.
Where can I find the official FDA record?
The official record is available on FDA.gov. Search for "Z-2342-2025" on FDA.gov for the full record. You can also use the openFDA API to query this record programmatically.
Want to know the moment something like this happens?
Get alerts for deviceRelated Actions
MAHURKAR 12 Fr High Pressure Triple Lumen Acute Dialysis Catheter, 24 cm, Straight Extensions, IC Tray REF: 888810200
Covidien LP · 2023-08-23
Sterile Procedural Trays, labeled as the following: a. BREAST AUGMENTATION b. BREAST REDUCTION CDS c. CUH ABDOMINO
MEDLINE INDUSTRIES, LP - Northfield · 2023-02-15
VTC Regular Kit Nephrostomy Catheter System Kit, Material Number REF M001245300; to provide external drainage of the uri
Boston Scientific Corporation · 2024-09-25
EnChroma Safety Glasses with Rx Indoor Universal Lenses: Martinez Cx1 Indoor Rx, SKU: Cx.PC.IN.MTZ.BK.Rx; Summit Indoo
Enchroma Inc · 2024-11-27
Otological Ventilation Tube - T-Tube 9mm - Silicone; Product Code: NZ3309;
Adept Medical Ltd · 2025-01-29