RecallHawk
Class II Recall

GE HealthCare OEC 9800.

GE Medical Systems, LLC

Summary

The FDA issued a Class II for GE HealthCare OEC 9800. by GE Medical Systems, LLC. Reason: GE HealthCare has become aware of a non-conformance to 21 CFR 1020.30 (k) in certain OEC 9800 and OEC 9900 Elite C-arm Systems. Specifically, impacted.

Details

Source

Device Recall

External ID

Z-2341-2025

Action Date

2025-09-03

Status

Ongoing

Category

device

Product Description

GE HealthCare OEC 9800.

Lot/Code Info: GTIN: 00840682114349

Reason for Recall

GE HealthCare has become aware of a non-conformance to 21 CFR 1020.30 (k) in certain OEC 9800 and OEC 9900 Elite C-arm Systems. Specifically, impacted systems are missing the full layer of lead tape in the collimator covers, which will lead to a small increase in radiation exposure to the patient and/or operator. The missing lead tape is not detectable by the user.

Distribution

US Nationwide distribution.

Type: FDA Mandated

Recall Initiated: 2025-07-02

Company

Context & Analysis

This is a Class II recall. Class II recalls indicate that use of or exposure to the product may cause temporary or medically reversible health consequences, or that the probability of serious consequences is remote.

This was one of 201 device recalls issued in the same week, part of 403 device-related FDA actions this month.

GE Medical Systems, LLC has 138 FDA actions in our database, including 121 recalls and 17 clearances.

Frequently Asked Questions

What is a Class II recall?

A Class II recall means the product may cause temporary or medically reversible adverse health consequences, or the probability of serious adverse consequences is remote. Consumers should discontinue use.

What should I do if I have this product?

Stop using the product. Check the product against the specific lot numbers or identifiers listed in the recall. Contact the manufacturer (GE Medical Systems, LLC) or your retailer for a refund or replacement. If you experienced adverse effects, report them to the FDA via MedWatch.

How often does GE Medical Systems, LLC have FDA actions?

GE Medical Systems, LLC has 138 FDA actions in our database, including 121 recalls and 17 device clearances.

Where can I find the official FDA record?

The official record is available on FDA.gov. Search for "Z-2341-2025" on FDA.gov for the full record. You can also use the openFDA API to query this record programmatically.