RecallHawk
Class II Recall

Stryker SmartPump Tourniquet, disposable sterile (single use) DISP 18X3,1BLA,1PRT QUICK / 5921-018-135; DISP 18X3

Stryker Corporation

Summary

The FDA issued a Class II for Stryker SmartPump Tourniquet, disposable sterile (single use) DISP 18X3,1BLA, by Stryker Corporation. Reason: Tourniquet cuff flange may become detached from the bladder. If leak occurs during the procedure, there may be risk of hemorrhage, resulting in blood .

Details

Source

Device Recall

External ID

Z-2277-2025

Action Date

2025-08-13

Status

Ongoing

Category

device

Product Description

Stryker SmartPump Tourniquet, disposable sterile (single use) DISP 18X3,1BLA,1PRT QUICK / 5921-018-135; DISP 18X3,1BLA,2PRT QUICK / 5921-018-235; DISP 24X4,1BLA, 2PRT QUICK / 5921-024-235

Lot/Code Info: 5921-018-135 GTIN 07613154599257 5921-018-235 GTIN 07613154599295 5921-024-235 GTIN 07613154599370

Quantity Affected: 233.230

Reason for Recall

Tourniquet cuff flange may become detached from the bladder. If leak occurs during the procedure, there may be risk of hemorrhage, resulting in blood loss greater than expected. Other serious risks include cerebral hypotension and cardiovascular collapse and arrest, risking stroke or death. Specific to IVRA, patient toxicity may also occur due to rapid systemic absorption of anesthetics.

Distribution

Domestic US distribution nationwide. International distribution to Australia Canada Hong Kong India Malaysia Netherlands New Zealand Singapore Sweden Thailand United Kingdom

Type: Voluntary: Firm initiated

Recall Initiated: 2025-07-02

Company

Context & Analysis

This is a Class II recall. Class II recalls indicate that use of or exposure to the product may cause temporary or medically reversible health consequences, or that the probability of serious consequences is remote.

This was one of 172 device recalls issued in the same week, part of 403 device-related FDA actions this month.

Stryker Corporation has 98 FDA actions in our database, including 90 recalls and 8 clearances.

Frequently Asked Questions

What is a Class II recall?

A Class II recall means the product may cause temporary or medically reversible adverse health consequences, or the probability of serious adverse consequences is remote. Consumers should discontinue use.

What should I do if I have this product?

Stop using the product. Check the product against the specific lot numbers or identifiers listed in the recall. Contact the manufacturer (Stryker Corporation) or your retailer for a refund or replacement. If you experienced adverse effects, report them to the FDA via MedWatch.

How often does Stryker Corporation have FDA actions?

Stryker Corporation has 98 FDA actions in our database, including 90 recalls and 8 device clearances.

Where can I find the official FDA record?

The official record is available on FDA.gov. Search for "Z-2277-2025" on FDA.gov for the full record. You can also use the openFDA API to query this record programmatically.