RecallHawk
Class II Recall

Brand Name: Exacta Mix Product Name: Vented Micro-Volume Inlet Model/Catalog Number: H938175 Product Description:

Baxter Healthcare Corporation

Summary

The FDA issued a Class II for Brand Name: Exacta Mix Product Name: Vented Micro-Volume Inlet Model/Catalog N by Baxter Healthcare Corporation. Reason: Affected lots may contain incorrect 0.8-micron filters. Use of a 0.8-micron filter in place of the validated 3-micron filter may lead to reduced flow .

Details

Source

Device Recall

External ID

Z-2262-2026

Action Date

2026-06-10

Status

Ongoing

Category

device

Product Description

Brand Name: Exacta Mix Product Name: Vented Micro-Volume Inlet Model/Catalog Number: H938175 Product Description: Tubing sets with push-on connector on one end (to attach to the Fluid Selector Valve) and a spike or other appropriate connector on the opposite end to attach the source container.

Lot/Code Info: Lot Code: Lot numbers: 804084, 804088, and 804089 UDI Number: 00085412475806

Quantity Affected: 105675

Reason for Recall

Affected lots may contain incorrect 0.8-micron filters. Use of a 0.8-micron filter in place of the validated 3-micron filter may lead to reduced flow efficiency, incomplete or slower transfers of ingredients, or system alarms. This issue may lead to clinical effects such as electrolyte imbalance, metabolic instability, delayed recovery, etc.

Distribution

Worldwide - US Nationwide distribution including in the states of Alabama, Arkansas, Arizona, California, Colorado, Florida, Georgia, Hawaii, Iowa, Idaho, Illinois, Indiana, Kansas, Kentucky, Louisiana, Massachusetts, Maine, Maryland, Michigan, Minnesota, Missouri, Mississippi, Montana, North Carolina, North Dakota, Nebraska, New Hampshire, New Jersey, New Mexico, Nevada, New York, Ohio, Oklahoma, Oregon, Pennsylvania, South Dakota, Tennessee, Texas, Utah, Virginia, Washington, Wisconsin, and West Virginia. The countries of Taiwan and Uruguay.

Type: Voluntary: Firm initiated

Recall Initiated: 2026-04-22

Company

Context & Analysis

This is a Class II recall. Class II recalls indicate that use of or exposure to the product may cause temporary or medically reversible health consequences, or that the probability of serious consequences is remote.

This was one of 169 device recalls issued in the same week, part of 403 device-related FDA actions this month.

Baxter Healthcare Corporation has 310 FDA actions in our database, including 286 recalls and 24 clearances.

Frequently Asked Questions

What is a Class II recall?

A Class II recall means the product may cause temporary or medically reversible adverse health consequences, or the probability of serious adverse consequences is remote. Consumers should discontinue use.

What should I do if I have this product?

Stop using the product. Check the product against the specific lot numbers or identifiers listed in the recall. Contact the manufacturer (Baxter Healthcare Corporation) or your retailer for a refund or replacement. If you experienced adverse effects, report them to the FDA via MedWatch.

How often does Baxter Healthcare Corporation have FDA actions?

Baxter Healthcare Corporation has 310 FDA actions in our database, including 286 recalls and 24 device clearances.

Where can I find the official FDA record?

The official record is available on FDA.gov. Search for "Z-2262-2026" on FDA.gov for the full record. You can also use the openFDA API to query this record programmatically.