RecallHawk
Class II Recall

Cios Select VA 21 with FD- A Mobile X-ray system intended for use in Operating room, Traumatology, Endoscopy, Intensive

Siemens Medical Solutions USA, Inc

Summary

The FDA issued a Class II for Cios Select VA 21 with FD- A Mobile X-ray system intended for use in Operating r by Siemens Medical Solutions USA, Inc. Reason: Incorrect glue was applied on the plexiglass sheet on the single tank cover assembly could potentially fall off and the ingress of liquid may occur, t.

Details

Source

Device Recall

External ID

Z-2240-2023

Action Date

2023-08-02

Status

Ongoing

Category

device

Product Description

Cios Select VA 21 with FD- A Mobile X-ray system intended for use in Operating room, Traumatology, Endoscopy, Intensive Care Station, Pediatrics, Ambulatory patient care and in Veterinary Medicine. Material #11515088

Lot/Code Info: UDI-DI :04056869148243 US S/N: 30001 30002 30008 30013 30029 30031 30032 30033 30034

Quantity Affected: 9 units

Reason for Recall

Incorrect glue was applied on the plexiglass sheet on the single tank cover assembly could potentially fall off and the ingress of liquid may occur, the risk of electric shock exists (e.g., ingress of a large amount of liquid following the plexiglass dropping off, at the same time the user or operator happens to touch the liquid on the cover which is connected to the internal live parts).

Distribution

Worldwide - US Nationwide distribution in the states of FL, GA, KS, MN, NC, NY and the countries of Austria, Bangladesh, Brazil, Canada, China, Croatia, El Salvador, France, Germany, Indonesia, Italy, Japan, Jordan, Lebanon, Luxembourg, Mauritius, Mexico, Monaco, Peru, Philippines, Poland, Romania, Slovakia, South Africa, Spain, Switzerland, Taiwan, Ukraine, United Kingdom.

Type: Voluntary: Firm initiated

Recall Initiated: 2023-06-05

Context & Analysis

This is a Class II recall. Class II recalls indicate that use of or exposure to the product may cause temporary or medically reversible health consequences, or that the probability of serious consequences is remote.

This was one of 191 device recalls issued in the same week, part of 413 device-related FDA actions this month.

Siemens Medical Solutions USA, Inc has 277 FDA actions in our database, including 173 recalls and 104 clearances.

Frequently Asked Questions

What is a Class II recall?

A Class II recall means the product may cause temporary or medically reversible adverse health consequences, or the probability of serious adverse consequences is remote. Consumers should discontinue use.

What should I do if I have this product?

Stop using the product. Check the product against the specific lot numbers or identifiers listed in the recall. Contact the manufacturer (Siemens Medical Solutions USA, Inc) or your retailer for a refund or replacement. If you experienced adverse effects, report them to the FDA via MedWatch.

How often does Siemens Medical Solutions USA, Inc have FDA actions?

Siemens Medical Solutions USA, Inc has 277 FDA actions in our database, including 173 recalls and 104 device clearances.

Where can I find the official FDA record?

The official record is available on FDA.gov. Search for "Z-2240-2023" on FDA.gov for the full record. You can also use the openFDA API to query this record programmatically.

Want to know the moment something like this happens?

Get alerts for device

Related Actions