ARTIS pheno, Model 10849000. interventional fluoroscopic x-ray system
Summary
The FDA issued a Class II for ARTIS pheno, Model 10849000. interventional fluoroscopic x-ray system by Siemens Medical Solutions USA, Inc. Reason: During patient examinations, the system may sporadically display a 0 x-ray dose administered to the patient. This behavior has only been observed du.
Details
Source
Device Recall
External ID
Z-2205-2026
Action Date
2026-06-10
Status
Ongoing
Category
device
Product Description
ARTIS pheno, Model 10849000. interventional fluoroscopic x-ray system
Lot/Code Info: UDI: 4056869046877
Reason for Recall
During patient examinations, the system may sporadically display a 0 x-ray dose administered to the patient. This behavior has only been observed during fluoroscopy and/or single-frame acquisitions. Despite the reported x-ray dose showing 0, an actual x-ray dose is applied to the patient.
Distribution
US and Worldwide
Type: FDA Mandated
Recall Initiated: 2026-03-09
Company
Malvern, PA
Context & Analysis
This is a Class II recall. Class II recalls indicate that use of or exposure to the product may cause temporary or medically reversible health consequences, or that the probability of serious consequences is remote.
This was one of 169 device recalls issued in the same week, part of 413 device-related FDA actions this month.
Siemens Medical Solutions USA, Inc has 277 FDA actions in our database, including 173 recalls and 104 clearances.
Frequently Asked Questions
What is a Class II recall?
A Class II recall means the product may cause temporary or medically reversible adverse health consequences, or the probability of serious adverse consequences is remote. Consumers should discontinue use.
What should I do if I have this product?
Stop using the product. Check the product against the specific lot numbers or identifiers listed in the recall. Contact the manufacturer (Siemens Medical Solutions USA, Inc) or your retailer for a refund or replacement. If you experienced adverse effects, report them to the FDA via MedWatch.
How often does Siemens Medical Solutions USA, Inc have FDA actions?
Siemens Medical Solutions USA, Inc has 277 FDA actions in our database, including 173 recalls and 104 device clearances.
Where can I find the official FDA record?
The official record is available on FDA.gov. Search for "Z-2205-2026" on FDA.gov for the full record. You can also use the openFDA API to query this record programmatically.
Want to know the moment something like this happens?
Get alerts for deviceRelated Actions
MAHURKAR 12 Fr High Pressure Triple Lumen Acute Dialysis Catheter, 24 cm, Straight Extensions, IC Tray REF: 888810200
Covidien LP · 2023-08-23
Sterile Procedural Trays, labeled as the following: a. BREAST AUGMENTATION b. BREAST REDUCTION CDS c. CUH ABDOMINO
MEDLINE INDUSTRIES, LP - Northfield · 2023-02-15
VTC Regular Kit Nephrostomy Catheter System Kit, Material Number REF M001245300; to provide external drainage of the uri
Boston Scientific Corporation · 2024-09-25
EnChroma Safety Glasses with Rx Indoor Universal Lenses: Martinez Cx1 Indoor Rx, SKU: Cx.PC.IN.MTZ.BK.Rx; Summit Indoo
Enchroma Inc · 2024-11-27
Otological Ventilation Tube - T-Tube 9mm - Silicone; Product Code: NZ3309;
Adept Medical Ltd · 2025-01-29