Accu-Chek Guide (SC) Kit -Intended to quantitatively measure glucose in fresh capillary whole blood from the fingertip,

Roche Diabetes Care, Inc.

Summary

The FDA issued a Class II for Accu-Chek Guide (SC) Kit -Intended to quantitatively measure glucose in fresh ca by Roche Diabetes Care, Inc.. Reason: The reason for recall is the meters may show an incorrect measurement unit (mmol/L rather than mg/dL) and may result in the wrong unit of measure. Wh.

Details

Source

Device Recall

External ID

Z-2174-2024

Action Date

2024-07-03

Status

Ongoing

Category

device

Product Description

Accu-Chek Guide (SC) Kit -Intended to quantitatively measure glucose in fresh capillary whole blood from the fingertip, palm, and upper arm as an aid in monitoring the effectiveness of glucose control. Product REF Number: 08453071001

Lot/Code Info: GTIN: 00365702729100 Serial Numbers: US: 92339920445, 92340117408, 92339920116, 92339955415, 92340120006, 92339094787, 92340116052,92340120057, 92339744998. Canada:93040305146

Quantity Affected: 48 units

Reason for Recall

The reason for recall is the meters may show an incorrect measurement unit (mmol/L rather than mg/dL) and may result in the wrong unit of measure. Which may subsequently cause an error in interpreting a result as approximately 18 times lower that the actual blood glucose. Potential immediate health consequences of the glucose meter providing a reading in the wrong units of measurement (UOM) may result in inappropriate rescue therapy for presumed hypoglycemia including urgent and repeat administration of carbohydrates. This in turn could lead to frequent episodes of hyperglycemia which is unrecognized and undertreated.Long range consequences of continued use of the affected meters could include mismanagement of the patient s diabetes for longer periods of time potentially resulting in a major health event (in general persistent and/or severe hyperglycemia) that may lead them to seek medical attention, and depending on the duration of hyperglycemia, could contribute to the development of microvascular (i.e., retinopathy, neuropathy, nephropathy) and/or macrovascular (i.e., myocardial infarction, ischemic stroke) complications of diabetes and could lead to fetal damage (abnormal fetal growth and altered organ development and maturation, fetal hypoxemia, spontaneous abortion, and congenital anomalies) in pregnant women. This issue was identified by three initial consumer complaints, with reports that the meters were presenting the incorrect unit of measure.

Distribution

Nationwide and Worldwide

Type: Voluntary: Firm initiated

Recall Initiated: 2024-05-13

Company

Context & Analysis

This is a Class II recall. Class II recalls indicate that use of or exposure to the product may cause temporary or medically reversible health consequences, or that the probability of serious consequences is remote.

This was one of 167 device recalls issued in the same week, part of 403 device-related FDA actions this month.

Roche Diabetes Care, Inc. has 13 FDA actions in our database, including 6 recalls and 7 clearances.

Frequently Asked Questions

What is a Class II recall?

A Class II recall means the product may cause temporary or medically reversible adverse health consequences, or the probability of serious adverse consequences is remote. Consumers should discontinue use.

What should I do if I have this product?

Stop using the product. Check the product against the specific lot numbers or identifiers listed in the recall. Contact the manufacturer (Roche Diabetes Care, Inc.) or your retailer for a refund or replacement. If you experienced adverse effects, report them to the FDA via MedWatch.

How often does Roche Diabetes Care, Inc. have FDA actions?

Roche Diabetes Care, Inc. has 13 FDA actions in our database, including 6 recalls and 7 device clearances.

Where can I find the official FDA record?

The official record is available on FDA.gov. Search for "Z-2174-2024" on FDA.gov for the full record. You can also use the openFDA API to query this record programmatically.

Want to know the moment something like this happens?

Get alerts for device

Related Actions