Medtronic CareLink SmartSync Device Manager, Model Number 24967A with the following software components: 1. Cobalt Cr

Medtronic, Inc.

Summary

The FDA issued a Class II for Medtronic CareLink SmartSync Device Manager, Model Number 24967A with the follow by Medtronic, Inc.. Reason: In prior SmartSync application versions, the Abort button stopped the test that was selected. During an induction test, there was a limited window of .

Details

Source

Device Recall

External ID

Z-2169-2025

Action Date

2025-07-30

Status

Ongoing

Category

device

Product Description

Medtronic CareLink SmartSync Device Manager, Model Number 24967A with the following software components: 1. Cobalt Crome application, Software Model Number D00U005; 2. Claria Amplia Compia application, Software Model Number D00U009; 3. Evera MRI application, Software Model Number D00U010; 3. Visia AF application, Software Model Number D00U011; 5. Viva Brava Evera application, Software Model Number D00U012;

Lot/Code Info: 1. Software Model Number, D00U003, GTIN 00763000002039, Updated Software Version 6.5.5; 2. Software Model Number, D00U004, GTIN 00763000002046, Updated Software Version 6.5.5; 3. Software Model Number, D00U005, GTIN 00763000002053, Updated Software Version 9.5.2; 4. Software Model Number, D00U006, GTIN 00763000397852, Updated Software Version 4.4.6; 5. Software Model Number, D00U007, GTIN 00763000397869, Updated Software Version 4.4.6; 6. Software Model Number, D00U008, GTIN 00763000397876, Updated Software Version 3.5.5; 7. Software Model Number, D00U009, GTIN 00763000397883, Updated Software Version 3.4.2; 8. Software Model Number, D00U010, GTIN 00763000397890, Updated Software Version 3.4.2; 9. Software Model Number, D00U011, GTIN 00763000397906, Updated Software Version 3.4.2; 10. Software Model Number, D00U012, GTIN 00763000397913, Updated Software Version 3.4.2; 11. Software Model Number, D00U022, GTIN 00763000544300, Updated Software Version 3.7.5.

Quantity Affected: 22091 total devices

Reason for Recall

In prior SmartSync application versions, the Abort button stopped the test that was selected. During an induction test, there was a limited window of time for the user to abort a therapy, thereby limiting the user's ability to cancel a high voltage therapy delivery.

Distribution

Worldwide - US Nationwide distribution.

Type: Voluntary: Firm initiated

Recall Initiated: 2025-06-18

Company

Medtronic, Inc.

Mounds View, MN

Context & Analysis

This is a Class II recall. Class II recalls indicate that use of or exposure to the product may cause temporary or medically reversible health consequences, or that the probability of serious consequences is remote.

This was one of 122 device recalls issued in the same week, part of 403 device-related FDA actions this month.

Medtronic, Inc. has 83 FDA actions in our database, including 46 recalls and 37 clearances.

Frequently Asked Questions

What is a Class II recall?

A Class II recall means the product may cause temporary or medically reversible adverse health consequences, or the probability of serious adverse consequences is remote. Consumers should discontinue use.

What should I do if I have this product?

Stop using the product. Check the product against the specific lot numbers or identifiers listed in the recall. Contact the manufacturer (Medtronic, Inc.) or your retailer for a refund or replacement. If you experienced adverse effects, report them to the FDA via MedWatch.

How often does Medtronic, Inc. have FDA actions?

Medtronic, Inc. has 83 FDA actions in our database, including 46 recalls and 37 device clearances.

Where can I find the official FDA record?

The official record is available on FDA.gov. Search for "Z-2169-2025" on FDA.gov for the full record. You can also use the openFDA API to query this record programmatically.

Want to know the moment something like this happens?

Get alerts for device

Related Actions