RecallHawk
Class II Recall

Estrone RIA, REF: DSL8700. Estrone RIA is an IVD used for the quantitative measurement of estrone in human serum and pl

Beckman Coulter Inc.

Summary

The FDA issued a Class II for Estrone RIA, REF: DSL8700. Estrone RIA is an IVD used for the quantitative meas by Beckman Coulter Inc.. Reason: A specific lot of bovine serum albumin (BSA) used for manufacturing of coated tubes (CT) that are included in the Estrone RIA kit lots caused differen.

Details

Source

Device Recall

External ID

Z-2152-2025

Action Date

2025-07-30

Status

Ongoing

Category

device

Product Description

Estrone RIA, REF: DSL8700. Estrone RIA is an IVD used for the quantitative measurement of estrone in human serum and plasma.

Lot/Code Info: UDI: 15099590211615/Lot: 250106D, 250203C, 250203D, 250303C, 250331C, 250428C, 250526C

Quantity Affected: 496 units

Reason for Recall

A specific lot of bovine serum albumin (BSA) used for manufacturing of coated tubes (CT) that are included in the Estrone RIA kit lots caused different affinity of manufactured coated tubes to patient samples, causing falsely increased assay results for the impacted Estrone RIA kits.

Distribution

US: North Carolina OUS: Argentina, Austria, Belgium, Canada, France, Germany, Greece, Hungary, Italy, Netherlands

Type: Voluntary: Firm initiated

Recall Initiated: 2025-06-12

Company

Context & Analysis

This is a Class II recall. Class II recalls indicate that use of or exposure to the product may cause temporary or medically reversible health consequences, or that the probability of serious consequences is remote.

This was one of 122 device recalls issued in the same week, part of 403 device-related FDA actions this month.

Beckman Coulter Inc. has 158 FDA actions in our database, including 111 recalls and 47 clearances.

Frequently Asked Questions

What is a Class II recall?

A Class II recall means the product may cause temporary or medically reversible adverse health consequences, or the probability of serious adverse consequences is remote. Consumers should discontinue use.

What should I do if I have this product?

Stop using the product. Check the product against the specific lot numbers or identifiers listed in the recall. Contact the manufacturer (Beckman Coulter Inc.) or your retailer for a refund or replacement. If you experienced adverse effects, report them to the FDA via MedWatch.

How often does Beckman Coulter Inc. have FDA actions?

Beckman Coulter Inc. has 158 FDA actions in our database, including 111 recalls and 47 device clearances.

Where can I find the official FDA record?

The official record is available on FDA.gov. Search for "Z-2152-2025" on FDA.gov for the full record. You can also use the openFDA API to query this record programmatically.