RecallHawk
Class II Recall

VITROS Chemistry Products CHE Slides (Japan); Catalog Numbers: (1) 684 4074 (90 slides), (2) 684 4138 (300 slides);

Ortho-Clinical Diagnostics, Inc.

Summary

The FDA issued a Class II for VITROS Chemistry Products CHE Slides (Japan); Catalog Numbers: (1) 684 4074 (90 by Ortho-Clinical Diagnostics, Inc.. Reason: Quality control (QC) drift when certain lots of Performance Verifiers (PVs) are used with certain lots of cholinesterase enzyme (CHE) slides. This dri.

Details

Source

Device Recall

External ID

Z-2123-2025

Action Date

2025-07-23

Status

Ongoing

Category

device

Product Description

VITROS Chemistry Products CHE Slides (Japan); Catalog Numbers: (1) 684 4074 (90 slides), (2) 684 4138 (300 slides);

Lot/Code Info: Catalog Numbers: (1) 684 4074 (90 slides), (2) 684 4138 (300 slides); UDI-DI: (1) 10758750030293, (2) 10758750030309; Lot Numbers: All lots within expiry.

Quantity Affected: 414,000 (OUS)

Reason for Recall

Quality control (QC) drift when certain lots of Performance Verifiers (PVs) are used with certain lots of cholinesterase enzyme (CHE) slides. This drift occurs when the PVs are used 2-7 days after reconstitution; however, the label stability claims for the PV fluids is 7 days following reconstitution. This may lead to inaccurate QC results. The use of the PVs when the QC results are outside of the customer s established ranges could result in delayed results for patients due to the need to repeat QC testing or reconstitute new PV fluids and repeat testing. It could also result in erroneous patient results (small positive bias).

Distribution

Worldwide Distribution: US (nationwide) to states of: AZ, CA, FL, GA, IL, MD, MI, MO, MS, NY, TN, TX, VA; and OUS (International) to countries of: Australia, Brazil, Canada, Chile, China, Colombia, Denmark, France, Germany, India, Italy, Japan, Mexico, Netherlands, Norway, Portugal, Russia, Singapore, Spain, Sweden, and United Kingdom.

Type: Voluntary: Firm initiated

Recall Initiated: 2025-06-10

Context & Analysis

This is a Class II recall. Class II recalls indicate that use of or exposure to the product may cause temporary or medically reversible health consequences, or that the probability of serious consequences is remote.

This was one of 109 device recalls issued in the same week, part of 403 device-related FDA actions this month.

Ortho-Clinical Diagnostics, Inc. has 71 FDA actions in our database, including 67 recalls and 4 clearances.

Frequently Asked Questions

What is a Class II recall?

A Class II recall means the product may cause temporary or medically reversible adverse health consequences, or the probability of serious adverse consequences is remote. Consumers should discontinue use.

What should I do if I have this product?

Stop using the product. Check the product against the specific lot numbers or identifiers listed in the recall. Contact the manufacturer (Ortho-Clinical Diagnostics, Inc.) or your retailer for a refund or replacement. If you experienced adverse effects, report them to the FDA via MedWatch.

How often does Ortho-Clinical Diagnostics, Inc. have FDA actions?

Ortho-Clinical Diagnostics, Inc. has 71 FDA actions in our database, including 67 recalls and 4 device clearances.

Where can I find the official FDA record?

The official record is available on FDA.gov. Search for "Z-2123-2025" on FDA.gov for the full record. You can also use the openFDA API to query this record programmatically.