Impella 5.5 with SmartAssist intravascular micro axial blood pump, Product Numbers 0550-0008 and 1000100

Abiomed, Inc.

Summary

The FDA issued a Class I for Impella 5.5 with SmartAssist intravascular micro axial blood pump, Product Numbe by Abiomed, Inc.. Reason: There is a potential risk for unintentional interaction of the Impella motor housing with the distal stent of a transcatheter aortic valve replacement.

Details

Source

Device Recall

External ID

Z-2114-2023

Action Date

2023-07-26

Status

Ongoing

Category

device

Product Description

Impella 5.5 with SmartAssist intravascular micro axial blood pump, Product Numbers 0550-0008 and 1000100

Lot/Code Info: UDI-DI: 00813502011531, 00813502012828;

Quantity Affected: 9252 units

Reason for Recall

There is a potential risk for unintentional interaction of the Impella motor housing with the distal stent of a transcatheter aortic valve replacement (TAVR) resulting in destruction of the impeller blades. This has resulted in low flow from the damaged Impella system. Systemic embolization of the fractured impeller material is a possibility.

Distribution

Worldwide distribution - US Nationwide.

Type: Voluntary: Firm initiated

Recall Initiated: 2023-06-14

Company

Abiomed, Inc.

Danvers, MA

Context & Analysis

This is a Class I recall. Class I recalls are the most serious — there is a reasonable probability that use of or exposure to the product will cause serious health consequences or death.

This was one of 192 device recalls issued in the same week, part of 413 device-related FDA actions this month.

Abiomed, Inc. has 66 FDA actions in our database, including 61 recalls and 5 clearances.

Frequently Asked Questions

What is a Class I recall?

A Class I recall is the most serious type. The FDA has determined there is a reasonable probability that the product will cause serious adverse health consequences or death. Consumers should stop using the product immediately.

What should I do if I have this product?

Stop using the product. Check the product against the specific lot numbers or identifiers listed in the recall. Contact the manufacturer (Abiomed, Inc.) or your retailer for a refund or replacement. If you experienced adverse effects, report them to the FDA via MedWatch.

How often does Abiomed, Inc. have FDA actions?

Abiomed, Inc. has 66 FDA actions in our database, including 61 recalls and 5 device clearances.

Where can I find the official FDA record?

The official record is available on FDA.gov. Search for "Z-2114-2023" on FDA.gov for the full record. You can also use the openFDA API to query this record programmatically.

Want to know the moment something like this happens?

Get alerts for device

Related Actions