RecallHawk
Class II Recall

VersaCross Transseptal Sheath Kit (VXS) Product Description/IDs VXS0201 VXS 63CM/45DEG/D0 SHEATH VXS2112 VERSACRO

Boston Scientific Corporation

Summary

The FDA issued a Class II for VersaCross Transseptal Sheath Kit (VXS) Product Description/IDs VXS0201 VXS 6 by Boston Scientific Corporation. Reason: Potential for hole in the Tyvek layer of the sterile barrier pouch, which may compromise device sterility. Issue could potentially lead to systemic in.

Details

Source

Device Recall

External ID

Z-2105-2025

Action Date

2025-07-16

Status

Ongoing

Category

device

Product Description

VersaCross Transseptal Sheath Kit (VXS) Product Description/IDs VXS0201 VXS 63CM/45DEG/D0 SHEATH VXS2112 VERSACROSS SHEATH (63CM 45DEG D0)-J Used for the percutaneous introduction of various types of cardiovascular catheters and guidewires to all heart chambers, including the left atrium via transseptal perforation / puncture.

Lot/Code Info: VXS0201 UDI-DI 00685447020301 Lot 35800622; VXS2112 UDI-DI 00685447019251 Lot 35892045

Quantity Affected: 14,724

Reason for Recall

Potential for hole in the Tyvek layer of the sterile barrier pouch, which may compromise device sterility. Issue could potentially lead to systemic infection and sepsis in worst case scenario. Potential punctures may go undetected in clinical settings due to small size and variable location.

Distribution

Worldwide distribution - US Nationwide and the countries of Canada, Hong Kong, and Japan.

Type: Voluntary: Firm initiated

Recall Initiated: 2025-06-10

Company

Context & Analysis

This is a Class II recall. Class II recalls indicate that use of or exposure to the product may cause temporary or medically reversible health consequences, or that the probability of serious consequences is remote.

This was one of 155 device recalls issued in the same week, part of 413 device-related FDA actions this month.

Boston Scientific Corporation has 399 FDA actions in our database, including 334 recalls and 65 clearances.

Frequently Asked Questions

What is a Class II recall?

A Class II recall means the product may cause temporary or medically reversible adverse health consequences, or the probability of serious adverse consequences is remote. Consumers should discontinue use.

What should I do if I have this product?

Stop using the product. Check the product against the specific lot numbers or identifiers listed in the recall. Contact the manufacturer (Boston Scientific Corporation) or your retailer for a refund or replacement. If you experienced adverse effects, report them to the FDA via MedWatch.

How often does Boston Scientific Corporation have FDA actions?

Boston Scientific Corporation has 399 FDA actions in our database, including 334 recalls and 65 device clearances.

Where can I find the official FDA record?

The official record is available on FDA.gov. Search for "Z-2105-2025" on FDA.gov for the full record. You can also use the openFDA API to query this record programmatically.

Want to know the moment something like this happens?

Get alerts for device

Related Actions