RecallHawk
Class II Recall

LinkSymphoKnee Distal Femoral Augment - Tilastan - Size. 7-8, Lateral-Right/Medial-Left, 5mm-Intended for primary and re

Waldemar Link GmbH & Co. KG (Mfg Site)

Summary

The FDA issued a Class II for LinkSymphoKnee Distal Femoral Augment - Tilastan - Size. 7-8, Lateral-Right/Medi by Waldemar Link GmbH & Co. KG (Mfg Site). Reason: Fixation screw was inserted through the wrong side of the augment (distal to proximal, as opposed to proximal to distal, which is the correct orientat.

Details

Source

Device Recall

External ID

Z-2067-2023

Action Date

2023-07-12

Status

Ongoing

Category

device

Product Description

LinkSymphoKnee Distal Femoral Augment - Tilastan - Size. 7-8, Lateral-Right/Medial-Left, 5mm-Intended for primary and revision total knee replacement in skeletally mature patients. Product Code: 880-307/21

Lot/Code Info: UDI-DI: 04026575257393 Lot Number: 2123233

Quantity Affected: 20 units

Reason for Recall

Fixation screw was inserted through the wrong side of the augment (distal to proximal, as opposed to proximal to distal, which is the correct orientation).

Distribution

US Nationwide distribution in the states of AL, IL, IN, LA, GA, NV, TX.

Type: Voluntary: Firm initiated

Recall Initiated: 2022-12-23

Context & Analysis

This is a Class II recall. Class II recalls indicate that use of or exposure to the product may cause temporary or medically reversible health consequences, or that the probability of serious consequences is remote.

This was one of 287 device recalls issued in the same week, part of 403 device-related FDA actions this month.

Waldemar Link GmbH & Co. KG (Mfg Site) has 61 FDA actions in our database, including 61 recalls.

Frequently Asked Questions

What is a Class II recall?

A Class II recall means the product may cause temporary or medically reversible adverse health consequences, or the probability of serious adverse consequences is remote. Consumers should discontinue use.

What should I do if I have this product?

Stop using the product. Check the product against the specific lot numbers or identifiers listed in the recall. Contact the manufacturer (Waldemar Link GmbH & Co. KG (Mfg Site)) or your retailer for a refund or replacement. If you experienced adverse effects, report them to the FDA via MedWatch.

How often does Waldemar Link GmbH & Co. KG (Mfg Site) have FDA actions?

Waldemar Link GmbH & Co. KG (Mfg Site) has 61 FDA actions in our database, including 61 recalls and 0 device clearances.

Where can I find the official FDA record?

The official record is available on FDA.gov. Search for "Z-2067-2023" on FDA.gov for the full record. You can also use the openFDA API to query this record programmatically.

Want to know the moment something like this happens?