RecallHawk
Class II Recall

Philips Tempus LS-Manual Defibrillator, Part Number: 00-3020 when using Tempus LS-Manual User/Operator Manual

Remote Diagnostic Technologies Ltd.

Summary

The FDA issued a Class II for Philips Tempus LS-Manual Defibrillator, Part Number: 00-3020 when using Tempus L by Remote Diagnostic Technologies Ltd.. Reason: Simultaneous ECG measurements with multiple devices, including a manual defibrillator, on the same patient may cause electrical noise or interference,.

Details

Source

Device Recall

External ID

Z-2028-2024

Action Date

2024-06-12

Status

Ongoing

Category

device

Product Description

Philips Tempus LS-Manual Defibrillator, Part Number: 00-3020 when using Tempus LS-Manual User/Operator Manual

Lot/Code Info: 00-3020: UDI-DI-07613365002737, Serial Numbers prior to 7022.002871 User/Operator Manual Versions: 41-3007EN-01 and prior

Quantity Affected: 1745

Reason for Recall

Simultaneous ECG measurements with multiple devices, including a manual defibrillator, on the same patient may cause electrical noise or interference, which may lead to distortions of the ECG signal; distortions may lead to no or ineffective pacing, or no or inappropriate defibrillation pulse.

Distribution

US Nationwide distribution including in the states of VA, DC, MA, WY, AZ, FL, CA, NM, NJ, TN, OH, KS, IN, NY, IL, TX, NV, NE, PA, AL, IA, GA, MO, KY, WI, OR, CO, WV, WA, ET, UT, ID, OK, MN, ME, MT, MI, NC, ND, MD, DE, VT.

Type: Voluntary: Firm initiated

Recall Initiated: 2024-05-09

Context & Analysis

This is a Class II recall. Class II recalls indicate that use of or exposure to the product may cause temporary or medically reversible health consequences, or that the probability of serious consequences is remote.

This was one of 212 device recalls issued in the same week, part of 403 device-related FDA actions this month.

Remote Diagnostic Technologies Ltd. has 11 FDA actions in our database, including 11 recalls.

Frequently Asked Questions

What is a Class II recall?

A Class II recall means the product may cause temporary or medically reversible adverse health consequences, or the probability of serious adverse consequences is remote. Consumers should discontinue use.

What should I do if I have this product?

Stop using the product. Check the product against the specific lot numbers or identifiers listed in the recall. Contact the manufacturer (Remote Diagnostic Technologies Ltd.) or your retailer for a refund or replacement. If you experienced adverse effects, report them to the FDA via MedWatch.

How often does Remote Diagnostic Technologies Ltd. have FDA actions?

Remote Diagnostic Technologies Ltd. has 11 FDA actions in our database, including 11 recalls and 0 device clearances.

Where can I find the official FDA record?

The official record is available on FDA.gov. Search for "Z-2028-2024" on FDA.gov for the full record. You can also use the openFDA API to query this record programmatically.

Want to know the moment something like this happens?

Get alerts for device

Related Actions