A-Series BiPAP Hybrid A30 (not marketed in US), A-Series BiPAP V30 Auto Continuous Ventilator, Minimum Ventilatory Suppo
Summary
The FDA issued a Class I for A-Series BiPAP Hybrid A30 (not marketed in US), A-Series BiPAP V30 Auto Continuo by Philips Respironics, Inc.. Reason: The polyester-based polyurethane (PE-PUR) foam may degrade into particles which may enter the devices air pathway and be ingested or inhaled by the us.
Details
Source
Device Recall
External ID
Z-1957-2021
Action Date
2021-07-21
Status
Ongoing
Category
device
Product Description
A-Series BiPAP Hybrid A30 (not marketed in US), A-Series BiPAP V30 Auto Continuous Ventilator, Minimum Ventilatory Support, Facility Use
Lot/Code Info: All Devices manufactured before 26 April 2021, All device serial numbers. Material Numbers 1111181 1111182 1111178 1140184 1135427 1143671
Quantity Affected: 8047
Reason for Recall
The polyester-based polyurethane (PE-PUR) foam may degrade into particles which may enter the devices air pathway and be ingested or inhaled by the user, and the PE-PUR foam may off-gas certain chemicals.
Distribution
Worldwide distribution including US Nationwide, Argentina, Australia, Brazil, Canada, China, France, Germany, Italy, Japan, Korea, and Latin America.
Type: Voluntary: Firm initiated
Recall Initiated: 2021-06-14
Company
Murrysville, PA
Context & Analysis
This is a Class I recall. Class I recalls are the most serious — there is a reasonable probability that use of or exposure to the product will cause serious health consequences or death.
This was one of 157 device recalls issued in the same week, part of 413 device-related FDA actions this month.
Philips Respironics, Inc. has 74 FDA actions in our database, including 74 recalls.
Frequently Asked Questions
What is a Class I recall?
A Class I recall is the most serious type. The FDA has determined there is a reasonable probability that the product will cause serious adverse health consequences or death. Consumers should stop using the product immediately.
What should I do if I have this product?
Stop using the product. Check the product against the specific lot numbers or identifiers listed in the recall. Contact the manufacturer (Philips Respironics, Inc.) or your retailer for a refund or replacement. If you experienced adverse effects, report them to the FDA via MedWatch.
How often does Philips Respironics, Inc. have FDA actions?
Philips Respironics, Inc. has 74 FDA actions in our database, including 74 recalls and 0 device clearances.
Where can I find the official FDA record?
The official record is available on FDA.gov. Search for "Z-1957-2021" on FDA.gov for the full record. You can also use the openFDA API to query this record programmatically.
Want to know the moment something like this happens?
Get alerts for deviceRelated Actions
MAHURKAR 12 Fr High Pressure Triple Lumen Acute Dialysis Catheter, 24 cm, Straight Extensions, IC Tray REF: 888810200
Covidien LP · 2023-08-23
Sterile Procedural Trays, labeled as the following: a. BREAST AUGMENTATION b. BREAST REDUCTION CDS c. CUH ABDOMINO
MEDLINE INDUSTRIES, LP - Northfield · 2023-02-15
VTC Regular Kit Nephrostomy Catheter System Kit, Material Number REF M001245300; to provide external drainage of the uri