RecallHawk
Class I Recall

Aurous Centimeter Sizing Catheter RPN N5.0-35-100-P-10S-PIG-CSC-20-01 GPN G47308 Sizing catheters are single-use, s

Cook Incorporated

Summary

The FDA issued a Class I for Aurous Centimeter Sizing Catheter RPN N5.0-35-100-P-10S-PIG-CSC-20-01 GPN G473 by Cook Incorporated. Reason: Marker bands may be at an increased risk of cracking/breakage. Potential adverse events that may result include increased procedural time, harms assoc.

Details

Source

Device Recall

External ID

Z-1935-2026

Action Date

2026-05-13

Status

Ongoing

Category

device

Product Description

Aurous Centimeter Sizing Catheter RPN N5.0-35-100-P-10S-PIG-CSC-20-01 GPN G47308 Sizing catheters are single-use, sterile catheters with gold marker bands. These catheters are for use in angiographic procedures and available in a variety of lengths (70 and 100 cm) and tip configurations (straight, PIG, and VCF).

Lot/Code Info: UDI-DI 00827002473089 Lots 16858459 16869145 16857138 16763324 16750208 16752733 16821053 16713191

Quantity Affected: 271

Reason for Recall

Marker bands may be at an increased risk of cracking/breakage. Potential adverse events that may result include increased procedural time, harms associated with device fragmentation/separation, and vessel injury. In a worst-case scenario, device fragmentation and separation may cause life-threatening harm or death.

Distribution

US distribution nationwide. OUS distribution to Brunei, Brazil, Canda, China, Colombia, Costa Rica, Hong Kong, Israel, India, Japan, South Korea, Macau, New Zealand, Singapore, El Salvador, Taiwan, Uruguay

Type: Voluntary: Firm initiated

Recall Initiated: 2026-04-02

Company

Cook Incorporated

Bloomington, IN

Context & Analysis

This is a Class I recall. Class I recalls are the most serious — there is a reasonable probability that use of or exposure to the product will cause serious health consequences or death.

This was one of 250 device recalls issued in the same week, part of 403 device-related FDA actions this month.

Cook Incorporated has 88 FDA actions in our database, including 85 recalls and 3 clearances.

Frequently Asked Questions

What is a Class I recall?

A Class I recall is the most serious type. The FDA has determined there is a reasonable probability that the product will cause serious adverse health consequences or death. Consumers should stop using the product immediately.

What should I do if I have this product?

Stop using the product. Check the product against the specific lot numbers or identifiers listed in the recall. Contact the manufacturer (Cook Incorporated) or your retailer for a refund or replacement. If you experienced adverse effects, report them to the FDA via MedWatch.

How often does Cook Incorporated have FDA actions?

Cook Incorporated has 88 FDA actions in our database, including 85 recalls and 3 device clearances.

Where can I find the official FDA record?

The official record is available on FDA.gov. Search for "Z-1935-2026" on FDA.gov for the full record. You can also use the openFDA API to query this record programmatically.