RecallHawk
Class II Recall

Brand Name: Powered Laser Surgical Instrument Product Name: SOLTIVE Premium SuperPulsed Laser System Model/Catalog Num

Olympus Corporation of the Americas

Summary

The FDA issued a Class II for Brand Name: Powered Laser Surgical Instrument Product Name: SOLTIVE Premium Sup by Olympus Corporation of the Americas. Reason: Potential for thermal incident including an unexpected occurrence of smoke, sparks, burning, or unsteady flames during clinical use. The fracturing of.

Details

Source

Device Recall

External ID

Z-1921-2025

Action Date

2025-06-18

Status

Ongoing

Category

device

Product Description

Brand Name: Powered Laser Surgical Instrument Product Name: SOLTIVE Premium SuperPulsed Laser System Model/Catalog Number: TFL-PLS, TFL-PLS-JA (Japan only); Software Version: N/A Product Description: SOLTIVE Laser Systems use technology to increase procedural efficiency and flexibility to ureteroscopy, PCNL, and potential soft tissue applications. Component: N/A

Lot/Code Info: All lot/serial numbers for the following products: TFL-PLS UDI: 00821925044111; TFL-PLS-JA (Japan only);

Quantity Affected: 1549 units (1025 US, 524 OUS)

Reason for Recall

Potential for thermal incident including an unexpected occurrence of smoke, sparks, burning, or unsteady flames during clinical use. The fracturing of the fiber is the cause of, or contributes to, these thermal incidents.

Distribution

Worldwide distribution - US Nationwide and the countries of AU, CA, CL, DE, HK, JP, SG.

Type: Voluntary: Firm initiated

Recall Initiated: 2025-05-06

Company

Context & Analysis

This is a Class II recall. Class II recalls indicate that use of or exposure to the product may cause temporary or medically reversible health consequences, or that the probability of serious consequences is remote.

This was one of 180 device recalls issued in the same week, part of 413 device-related FDA actions this month.

Olympus Corporation of the Americas has 334 FDA actions in our database, including 334 recalls.

Frequently Asked Questions

What is a Class II recall?

A Class II recall means the product may cause temporary or medically reversible adverse health consequences, or the probability of serious adverse consequences is remote. Consumers should discontinue use.

What should I do if I have this product?

Stop using the product. Check the product against the specific lot numbers or identifiers listed in the recall. Contact the manufacturer (Olympus Corporation of the Americas) or your retailer for a refund or replacement. If you experienced adverse effects, report them to the FDA via MedWatch.

How often does Olympus Corporation of the Americas have FDA actions?

Olympus Corporation of the Americas has 334 FDA actions in our database, including 334 recalls and 0 device clearances.

Where can I find the official FDA record?

The official record is available on FDA.gov. Search for "Z-1921-2025" on FDA.gov for the full record. You can also use the openFDA API to query this record programmatically.

Want to know the moment something like this happens?

Get alerts for device

Related Actions