RecallHawk
Class II Recall

Signature Laparoscopic Instruments, Dolphin Nose Grasper with Rotating Shaft and Monopolar Electrosurgical Footswitch Co

Silex Medical, LLC

Summary

The FDA issued a Class II for Signature Laparoscopic Instruments, Dolphin Nose Grasper with Rotating Shaft and by Silex Medical, LLC. Reason: Insulation on the tube did not extend fully to the proximal end of the shaft on the instrument, leaving an exposed (not insulated) area of the shaft c.

Details

Source

Device Recall

External ID

Z-1893-2025

Action Date

2025-06-11

Status

Ongoing

Category

device

Product Description

Signature Laparoscopic Instruments, Dolphin Nose Grasper with Rotating Shaft and Monopolar Electrosurgical Footswitch Connector. Model/REF Number: 5860020.

Lot/Code Info: Model/REF Number: 5860020. UDI-DI: B33158600200. Lot Number: M28181. Serial Numbers: 0691, 0692.

Quantity Affected: 2 units

Reason for Recall

Insulation on the tube did not extend fully to the proximal end of the shaft on the instrument, leaving an exposed (not insulated) area of the shaft closer to the rotator wheel.

Distribution

US Nationwide distribution in the states of Florida, North Dakota, Illinois, California, Texas, Virginia, Michigan, Connecticut, Oregon, Iowa, Mississippi, New Jersey, Idaho, West Virginia, Hawaii.

Type: Voluntary: Firm initiated

Recall Initiated: 2024-01-09

Company

Silex Medical, LLC

Southington, CT

Context & Analysis

This is a Class II recall. Class II recalls indicate that use of or exposure to the product may cause temporary or medically reversible health consequences, or that the probability of serious consequences is remote.

This was one of 165 device recalls issued in the same week, part of 403 device-related FDA actions this month.

Silex Medical, LLC has 8 FDA actions in our database, including 8 recalls.

Frequently Asked Questions

What is a Class II recall?

A Class II recall means the product may cause temporary or medically reversible adverse health consequences, or the probability of serious adverse consequences is remote. Consumers should discontinue use.

What should I do if I have this product?

Stop using the product. Check the product against the specific lot numbers or identifiers listed in the recall. Contact the manufacturer (Silex Medical, LLC) or your retailer for a refund or replacement. If you experienced adverse effects, report them to the FDA via MedWatch.

How often does Silex Medical, LLC have FDA actions?

Silex Medical, LLC has 8 FDA actions in our database, including 8 recalls and 0 device clearances.

Where can I find the official FDA record?

The official record is available on FDA.gov. Search for "Z-1893-2025" on FDA.gov for the full record. You can also use the openFDA API to query this record programmatically.

Want to know the moment something like this happens?

Get alerts for device

Related Actions