MEDLINE CLEARPRO TRACH T-PIECE CLOSED SUCTION CATHETER, flat peel pouch EA, 10 EA per Inner Carton (IC), 20 EA/2 Inner c
Summary
The FDA issued a Class II for MEDLINE CLEARPRO TRACH T-PIECE CLOSED SUCTION CATHETER, flat peel pouch EA, 10 E by MEDLINE INDUSTRIES, LP - Northfield. Reason: Potential for the suction catheter to come apart from the green connector near the thumb valve. The connector is coming apart from the assembled devic.
Details
Source
Device Recall
External ID
Z-1857-2022
Action Date
2022-10-05
Status
Ongoing
Category
device
Product Description
MEDLINE CLEARPRO TRACH T-PIECE CLOSED SUCTION CATHETER, flat peel pouch EA, 10 EA per Inner Carton (IC), 20 EA/2 Inner cartons per Case (CS). Labeled as follows: a. TRACH T-PIECE, 14 Fr, Item Number/REF: DYNCPTP14; b. TRACH T-PIECE, 14 Fr, Item Number/REF: DYNCPTP14T
Lot/Code Info: a. UDI-DI (GTIN) case: (01)20888277703466; UDI-DI (GTIN) Each: (01)10888277703469, Lot Numbers 6922030011; b. UDI-DI (GTIN) case: (01)20888277703473; UDI-DI (GTIN) Each: (01)10888277703476, Lot Numbers 6921050011, 6921060011, 6921040012, 6921110011, 6921120011 6921120012, 6922010011, 6922010012, 6922010031, 6922020031, 6922030011.
Quantity Affected: 136,080 units
Reason for Recall
Potential for the suction catheter to come apart from the green connector near the thumb valve. The connector is coming apart from the assembled device when pulling the catheter out of the artificial airway in the patient suctioning process.
Distribution
Nationwide distribution to AK, AL, AR, AZ, CA, CO, CT, DC, FL, GA, HI, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, MT, NC, ND, NE, NH, NJ, NV, NY, OH, OK, OR, PA, RI, SC, SD, TN, TX, UT, VA, VT, WA, WI, WV, WY, and Puerto Rico. International distribution to Canada, Chile, Colombia, Panama, Qatar, United Arab Emirates
Type: Voluntary: Firm initiated
Recall Initiated: 2022-08-03
Company
Northfield, IL
Context & Analysis
This is a Class II recall. Class II recalls indicate that use of or exposure to the product may cause temporary or medically reversible health consequences, or that the probability of serious consequences is remote.
This was one of 185 device recalls issued in the same week, part of 413 device-related FDA actions this month.
MEDLINE INDUSTRIES, LP - Northfield has 822 FDA actions in our database, including 822 recalls.
Frequently Asked Questions
What is a Class II recall?
A Class II recall means the product may cause temporary or medically reversible adverse health consequences, or the probability of serious adverse consequences is remote. Consumers should discontinue use.
What should I do if I have this product?
Stop using the product. Check the product against the specific lot numbers or identifiers listed in the recall. Contact the manufacturer (MEDLINE INDUSTRIES, LP - Northfield) or your retailer for a refund or replacement. If you experienced adverse effects, report them to the FDA via MedWatch.
How often does MEDLINE INDUSTRIES, LP - Northfield have FDA actions?
MEDLINE INDUSTRIES, LP - Northfield has 822 FDA actions in our database, including 822 recalls and 0 device clearances.
Where can I find the official FDA record?
The official record is available on FDA.gov. Search for "Z-1857-2022" on FDA.gov for the full record. You can also use the openFDA API to query this record programmatically.
Want to know the moment something like this happens?
Get alerts for deviceRelated Actions
MAHURKAR 12 Fr High Pressure Triple Lumen Acute Dialysis Catheter, 24 cm, Straight Extensions, IC Tray REF: 888810200
Covidien LP · 2023-08-23
Sterile Procedural Trays, labeled as the following: a. BREAST AUGMENTATION b. BREAST REDUCTION CDS c. CUH ABDOMINO
MEDLINE INDUSTRIES, LP - Northfield · 2023-02-15
VTC Regular Kit Nephrostomy Catheter System Kit, Material Number REF M001245300; to provide external drainage of the uri
Boston Scientific Corporation · 2024-09-25
EnChroma Safety Glasses with Rx Indoor Universal Lenses: Martinez Cx1 Indoor Rx, SKU: Cx.PC.IN.MTZ.BK.Rx; Summit Indoo
Enchroma Inc · 2024-11-27
Otological Ventilation Tube - T-Tube 9mm - Silicone; Product Code: NZ3309;
Adept Medical Ltd · 2025-01-29