RecallHawk
Class II Recall

Airvo 2 Humidifier, REF: PT101US, and myAirvo 2 Humidifier, REF: PT100US

Fisher & Paykel Healthcare, Ltd.

Summary

The FDA issued a Class II for Airvo 2 Humidifier, REF: PT101US, and myAirvo 2 Humidifier, REF: PT100US by Fisher & Paykel Healthcare, Ltd.. Reason: Humidifier devices, used to deliver high flow respiratory therapy to patients, manufactured before 14 August 2017, have a speaker configuration that m.

Details

Source

Device Recall

External ID

Z-1823-2024

Action Date

2024-05-22

Status

Ongoing

Category

device

Product Description

Airvo 2 Humidifier, REF: PT101US, and myAirvo 2 Humidifier, REF: PT100US

Lot/Code Info: REF/UDI-DI: PT100US/09420012422248, PT101US 09420012422347. Serial Number Range: 120521XXXXXX - 170813XXXXXX Note: X digits are variable and do not affect the identification of the affected range. Devices manufactured before 14 August 2017

Quantity Affected: 7,147

Reason for Recall

Humidifier devices, used to deliver high flow respiratory therapy to patients, manufactured before 14 August 2017, have a speaker configuration that may result in distorted, intermittent or inaudible alarm sound levels. In the absence of an audible alarm, if there is an interruption to therapy, a patient may experience oxygen desaturation.

Distribution

US Nationwide distribution including in the states of AK, AL, AR, AZ, CA, CO, CT, DE, FL, GA, HI, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, MT, NC, ND, NE, NH, NJ, NM, NV, NY, OH, OK, OR, PA, RI, SC, SD, TN, TX, UT, VA, VT, WA, WI, WV, WY.

Type: Voluntary: Firm initiated

Recall Initiated: 2024-04-02

Context & Analysis

This is a Class II recall. Class II recalls indicate that use of or exposure to the product may cause temporary or medically reversible health consequences, or that the probability of serious consequences is remote.

This was one of 191 device recalls issued in the same week, part of 403 device-related FDA actions this month.

Fisher & Paykel Healthcare, Ltd. has 18 FDA actions in our database, including 3 recalls and 15 clearances.

Frequently Asked Questions

What is a Class II recall?

A Class II recall means the product may cause temporary or medically reversible adverse health consequences, or the probability of serious adverse consequences is remote. Consumers should discontinue use.

What should I do if I have this product?

Stop using the product. Check the product against the specific lot numbers or identifiers listed in the recall. Contact the manufacturer (Fisher & Paykel Healthcare, Ltd.) or your retailer for a refund or replacement. If you experienced adverse effects, report them to the FDA via MedWatch.

How often does Fisher & Paykel Healthcare, Ltd. have FDA actions?

Fisher & Paykel Healthcare, Ltd. has 18 FDA actions in our database, including 3 recalls and 15 device clearances.