RecallHawk
Class I Recall

B Braun Hemodialysis Bloodlines, LOW VOL, NO CHAMBER 4.8MM PUMP SEGMENT. Model Number: B3-4630M4705.

B Braun Medical Inc

Summary

The FDA issued a Class I for B Braun Hemodialysis Bloodlines, LOW VOL, NO CHAMBER 4.8MM PUMP SEGMENT. Model N by B Braun Medical Inc. Reason: Potential for the accumulation of small air bubbles in the arterial line due to adherence of blood gases to the tubing under negative pressure..

Details

Source

Device Recall

External ID

Z-1803-2026

Action Date

2026-04-22

Status

Ongoing

Category

device

Product Description

B Braun Hemodialysis Bloodlines, LOW VOL, NO CHAMBER 4.8MM PUMP SEGMENT. Model Number: B3-4630M4705.

Lot/Code Info: Model Number: B3-4630M4705; UDI-DI Primary: 04046955348848; UDI-DI Unit: 04046955348831; All lots manufactured since 05AUG2025.

Quantity Affected: 4,884 units

Reason for Recall

Potential for the accumulation of small air bubbles in the arterial line due to adherence of blood gases to the tubing under negative pressure.

Distribution

Worldwide distribution - US Nationwide and the country of Canada.

Type: Voluntary: Firm initiated

Recall Initiated: 2026-03-19

Company

B Braun Medical Inc

Bethlehem, PA

Context & Analysis

This is a Class I recall. Class I recalls are the most serious — there is a reasonable probability that use of or exposure to the product will cause serious health consequences or death.

This was one of 143 device recalls issued in the same week, part of 403 device-related FDA actions this month.

B Braun Medical Inc has 282 FDA actions in our database, including 269 recalls and 13 clearances.

Frequently Asked Questions

What is a Class I recall?

A Class I recall is the most serious type. The FDA has determined there is a reasonable probability that the product will cause serious adverse health consequences or death. Consumers should stop using the product immediately.

What should I do if I have this product?

Stop using the product. Check the product against the specific lot numbers or identifiers listed in the recall. Contact the manufacturer (B Braun Medical Inc) or your retailer for a refund or replacement. If you experienced adverse effects, report them to the FDA via MedWatch.

How often does B Braun Medical Inc have FDA actions?

B Braun Medical Inc has 282 FDA actions in our database, including 269 recalls and 13 device clearances.

Where can I find the official FDA record?

The official record is available on FDA.gov. Search for "Z-1803-2026" on FDA.gov for the full record. You can also use the openFDA API to query this record programmatically.

Want to know the moment something like this happens?

Get alerts for device

Related Actions