RecallHawk
Class II Recall

The Randox Serology I Positive Control contains the following analytes in human based plasma: Anti HIV 1/2, Anti HTLV 1/

Randox Laboratories Ltd.

Summary

The FDA issued a Class II for The Randox Serology I Positive Control contains the following analytes in human by Randox Laboratories Ltd.. Reason: ***Update** :HBsAg analyte is also Non-Reactive on the Ortho Vitros analyser as well as the Beckman Coulter DxI. Randox Serology I Positive Control w.

Details

Source

Device Recall

External ID

Z-1772-2022

Action Date

2022-09-28

Status

Terminated

Category

device

Product Description

The Randox Serology I Positive Control contains the following analytes in human based plasma: Anti HIV 1/2, Anti HTLV 1/2, Anti HCV, HBsAg, Anti HBc, and Anti Treponema pallidum IgG. The product is delivered as a liquid in a kit of 3 identical ready-to-use vials of 5 mL each.

Lot/Code Info: GTIN: 05055273216509 Lot number: 036SR Exp Date: 28th March 2023

Quantity Affected: 2 units

Reason for Recall

***Update** :HBsAg analyte is also Non-Reactive on the Ortho Vitros analyser as well as the Beckman Coulter DxI. Randox Serology I Positive Control will test negative (Non-Reactive) for Marker HBsAg resulting in delay in reporting patient results due to the positive control O36SR testing negative (Non-Reactive) for Marker HBsAg

Distribution

US Nationwide distribution in the states of WV, NC.

Type: Voluntary: Firm initiated

Recall Initiated: 2022-08-01

Company

Randox Laboratories Ltd.

Crumlin (North), N/A

Context & Analysis

This is a Class II recall. Class II recalls indicate that use of or exposure to the product may cause temporary or medically reversible health consequences, or that the probability of serious consequences is remote.

This was one of 145 device recalls issued in the same week, part of 403 device-related FDA actions this month.

Randox Laboratories Ltd. has 39 FDA actions in our database, including 37 recalls and 2 clearances.

Frequently Asked Questions

What is a Class II recall?

A Class II recall means the product may cause temporary or medically reversible adverse health consequences, or the probability of serious adverse consequences is remote. Consumers should discontinue use.

What should I do if I have this product?

Stop using the product. Check the product against the specific lot numbers or identifiers listed in the recall. Contact the manufacturer (Randox Laboratories Ltd.) or your retailer for a refund or replacement. If you experienced adverse effects, report them to the FDA via MedWatch.

How often does Randox Laboratories Ltd. have FDA actions?

Randox Laboratories Ltd. has 39 FDA actions in our database, including 37 recalls and 2 device clearances.

Where can I find the official FDA record?

The official record is available on FDA.gov. Search for "Z-1772-2022" on FDA.gov for the full record. You can also use the openFDA API to query this record programmatically.