Surpria 64: Software Version V3.11, V3.22

FUJIFILM Healthcare Americas Corporation

Summary

The FDA issued a Class II for Surpria 64: Software Version V3.11, V3.22 by FUJIFILM Healthcare Americas Corporation. Reason: Error 00003050 or 00003052 may occur during scanning, the captured images will not be reconstructed and the Raw Data will not be displayed in the wor.

Details

Source

Device Recall

External ID

Z-1750-2022

Action Date

2022-09-21

Status

Ongoing

Category

device

Product Description

Surpria 64: Software Version V3.11, V3.22

Lot/Code Info: UDI: (01)04573596200520(11)220215(21)W5086(250)G1Z0211021 (01)04580292766601(11)200130(21)W8013(250)G1Z0098019 (01)04580292766601(11)200226(21)W8014(250)G1Z0130719 (01)04580292766601(11)201217(21)W8015(250)G1Z0024020 (01)04580292766601(11)210217(21)W8016(250)G1Z0216720 (01)04580292766601(11)210528(21)W8017(250)G1Z0239620 (01)04580292766601(11)210610(21)W8018(250)G1Z0039321 (01)04580292766601(11)210628(21)W8019(250)G1Z0223420 (01)04573596200520(11)210727(21)W8020(250)G1Z0087821 (01)04573596200520(11)210825(21)W8021(250)G1Z0139021 (01)04573596200520(11)211012(21)W8022(250)G1Z0112721 (01)04573596200520(11)211026(21)W8023(250)G1Z0156921 (01)04573596200520(11)220113(21)W8024(250)G1Z0043121 (01)04573596200520(11)220113(21)W8025(250)G1Z0240221 (01)04573596200520(11)220204(21)W8026(250)G1Z0244021 (01)04573596200520(11)220204(21)W8027(250)G1Z0249821 (01)04573596200520(11)220214(21)W8028(250)G1Z0249921 Serial Numbers: G1Z0098019 G1Z0130719 G1Z0024020 G1Z0216720 G1Z0239620 G1Z0039321 G1Z0223420 G1Z0087821 G1Z0139021 G1Z0112721 G1Z0156921 G1Z0043121 G1Z0240221 G1Z0244021 G1Z0249821 G1Z0249921

Quantity Affected: 16 units

Reason for Recall

Error 00003050 or 00003052 may occur during scanning, the captured images will not be reconstructed and the Raw Data will not be displayed in the work list.result, may result in the need for additional scanning to capture the images properly for reconstruction, thereby exposing the patient to additional radiation

Distribution

US Nationwide Distribution

Type: Voluntary: Firm initiated

Recall Initiated: 2022-07-12

Context & Analysis

This is a Class II recall. Class II recalls indicate that use of or exposure to the product may cause temporary or medically reversible health consequences, or that the probability of serious consequences is remote.

This was one of 161 device recalls issued in the same week, part of 403 device-related FDA actions this month.

FUJIFILM Healthcare Americas Corporation has 36 FDA actions in our database, including 23 recalls and 13 clearances.

Frequently Asked Questions

What is a Class II recall?

A Class II recall means the product may cause temporary or medically reversible adverse health consequences, or the probability of serious adverse consequences is remote. Consumers should discontinue use.

What should I do if I have this product?

Stop using the product. Check the product against the specific lot numbers or identifiers listed in the recall. Contact the manufacturer (FUJIFILM Healthcare Americas Corporation) or your retailer for a refund or replacement. If you experienced adverse effects, report them to the FDA via MedWatch.

How often does FUJIFILM Healthcare Americas Corporation have FDA actions?

FUJIFILM Healthcare Americas Corporation has 36 FDA actions in our database, including 23 recalls and 13 device clearances.

Where can I find the official FDA record?

The official record is available on FDA.gov. Search for "Z-1750-2022" on FDA.gov for the full record. You can also use the openFDA API to query this record programmatically.

Want to know the moment something like this happens?

Get alerts for device

Related Actions