Philips Tempus Pro Patient Monitor, REF: 00-1004-R, 00-1007-R, 00-1024-R, 00-1026-R, 00-1004, 00-1007
Summary
The FDA issued a Class II for Philips Tempus Pro Patient Monitor, REF: 00-1004-R, 00-1007-R, 00-1024-R, 00-10 by Remote Diagnostic Technologies Ltd.. Reason: Patient Monitor software update to address issues:1)May freeze at startup with MPM Application error and Attention Restart Required messages,2)Monitor.
Details
Source
Device Recall
External ID
Z-1729-2025
Action Date
2025-05-14
Status
Ongoing
Category
device
Product Description
Philips Tempus Pro Patient Monitor, REF: 00-1004-R, 00-1007-R, 00-1024-R, 00-1026-R, 00-1004, 00-1007
Lot/Code Info: Software versions prior to vx.40. REF/UDI-DI: 00-1004-R/05060472440020, 05060472442901; 00-1007-R/05060472440013, 05060472442918; 00-1024-R/05060472441027, 05060472442925; 00-1026-R/05060472441058, 05060472442932; 00-1004/05060472440624; 00-1007/05060472440655
Quantity Affected: 6961
Reason for Recall
Patient Monitor software update to address issues:1)May freeze at startup with MPM Application error and Attention Restart Required messages,2)Monitor may freeze and reboot at start of 12-lead ECG or after, 3)May reboot after intermittent connection with Smart Mount, 4) 12-lead ECG measurements with Louvain algorithm and software before vx.36 could produce inaccurate rhythm/morphology statements.
Distribution
Worldwide - US Nationwide distribution including in the states of AZ, VA, FL, CA, TX, NM, IA, NJ, TN, OH, KS, IN, NC, IL, WY, WI, NV, NY, NE, PA, AL, , MA, GA, MO, CO, OK, KY, OR, MS, MI, HI, WA, UT, WV, ID, ND, NH, MN, MT, ME, MD, DE, DC and the countries of AU, GB, IT, SA, DE, AE, BO, FR, ES, BE, IL, SG, CA, IE, TH, NL, MY, AT, DK, RO, CH, IS, PT, NO, HK, PL, KY, OM, CZ, SI, SL, ZA, AF, SE, FI.
Type: Voluntary: Firm initiated
Recall Initiated: 2025-04-15
Company
Farnborough, N/A
Context & Analysis
This is a Class II recall. Class II recalls indicate that use of or exposure to the product may cause temporary or medically reversible health consequences, or that the probability of serious consequences is remote.
This was one of 146 device recalls issued in the same week, part of 403 device-related FDA actions this month.
Remote Diagnostic Technologies Ltd. has 11 FDA actions in our database, including 11 recalls.
Frequently Asked Questions
What is a Class II recall?
A Class II recall means the product may cause temporary or medically reversible adverse health consequences, or the probability of serious adverse consequences is remote. Consumers should discontinue use.
What should I do if I have this product?
Stop using the product. Check the product against the specific lot numbers or identifiers listed in the recall. Contact the manufacturer (Remote Diagnostic Technologies Ltd.) or your retailer for a refund or replacement. If you experienced adverse effects, report them to the FDA via MedWatch.
How often does Remote Diagnostic Technologies Ltd. have FDA actions?
Remote Diagnostic Technologies Ltd. has 11 FDA actions in our database, including 11 recalls and 0 device clearances.
Where can I find the official FDA record?
The official record is available on FDA.gov. Search for "Z-1729-2025" on FDA.gov for the full record. You can also use the openFDA API to query this record programmatically.
Want to know the moment something like this happens?
Get alerts for deviceRelated Actions
MAHURKAR 12 Fr High Pressure Triple Lumen Acute Dialysis Catheter, 24 cm, Straight Extensions, IC Tray REF: 888810200
Covidien LP · 2023-08-23
Sterile Procedural Trays, labeled as the following: a. BREAST AUGMENTATION b. BREAST REDUCTION CDS c. CUH ABDOMINO
MEDLINE INDUSTRIES, LP - Northfield · 2023-02-15
VTC Regular Kit Nephrostomy Catheter System Kit, Material Number REF M001245300; to provide external drainage of the uri
Boston Scientific Corporation · 2024-09-25
EnChroma Safety Glasses with Rx Indoor Universal Lenses: Martinez Cx1 Indoor Rx, SKU: Cx.PC.IN.MTZ.BK.Rx; Summit Indoo
Enchroma Inc · 2024-11-27
Otological Ventilation Tube - T-Tube 9mm - Silicone; Product Code: NZ3309;
Adept Medical Ltd · 2025-01-29