ZPLP (Zimmer Periarticular Locking Plate System) Distal Lateral Fibular Plate, Left, 8 Holes, 132 MM, Sterile- Indicated
Summary
The FDA issued a Class II for ZPLP (Zimmer Periarticular Locking Plate System) Distal Lateral Fibular Plate, L by Zimmer, Inc.. Reason: Thread form issue for the locking holes in that the locking screws may not properly mate with the plate. The improperly mated screw may not be readily.
Details
Source
Device Recall
External ID
Z-1717-2024
Action Date
2024-05-08
Status
Ongoing
Category
device
Product Description
ZPLP (Zimmer Periarticular Locking Plate System) Distal Lateral Fibular Plate, Left, 8 Holes, 132 MM, Sterile- Indicated for temporary internal fixation and stabilization of osteotomies and fractures Item Number: 47-2357-018-08
Lot/Code Info: UDI: (01)00889024088580(17)300112(10)64568169 (01)00889024088580(17)300307(10)64697763 (01)00889024088580(17)300225(10)64697764 (01)00889024088580(17)330215(10)65755769 (01)00889024088580(17)330218(10)65782566 (01)00889024088580(17)330211(10)65782567 (01)00889024088580(17)331208(10)66233085 Lot Numbers: 64568169 64697763 64697764 65755769 65782566 65782567 66233085
Reason for Recall
Thread form issue for the locking holes in that the locking screws may not properly mate with the plate. The improperly mated screw may not be readily recognizable by the user since the screw may not correctly lock, may result in delay in surgery or loss of fixation resulting further intervention
Distribution
Worldwide distribution - US Nationwide and the countries of Brazil, Canada, China, Hong Kong, India, Japan, New Zealand, Seoul Korea, Singapore, Switzerland, Taiwan.
Type: Voluntary: Firm initiated
Recall Initiated: 2024-03-20
Company
Warsaw, IN
Context & Analysis
This is a Class II recall. Class II recalls indicate that use of or exposure to the product may cause temporary or medically reversible health consequences, or that the probability of serious consequences is remote.
This was one of 175 device recalls issued in the same week, part of 403 device-related FDA actions this month.
Zimmer, Inc. has 87 FDA actions in our database, including 65 recalls and 22 clearances.
Frequently Asked Questions
What is a Class II recall?
A Class II recall means the product may cause temporary or medically reversible adverse health consequences, or the probability of serious adverse consequences is remote. Consumers should discontinue use.
What should I do if I have this product?
Stop using the product. Check the product against the specific lot numbers or identifiers listed in the recall. Contact the manufacturer (Zimmer, Inc.) or your retailer for a refund or replacement. If you experienced adverse effects, report them to the FDA via MedWatch.
How often does Zimmer, Inc. have FDA actions?
Zimmer, Inc. has 87 FDA actions in our database, including 65 recalls and 22 device clearances.
Where can I find the official FDA record?
The official record is available on FDA.gov. Search for "Z-1717-2024" on FDA.gov for the full record. You can also use the openFDA API to query this record programmatically.
Want to know the moment something like this happens?
Get alerts for deviceRelated Actions
MAHURKAR 12 Fr High Pressure Triple Lumen Acute Dialysis Catheter, 24 cm, Straight Extensions, IC Tray REF: 888810200
Covidien LP · 2023-08-23
Sterile Procedural Trays, labeled as the following: a. BREAST AUGMENTATION b. BREAST REDUCTION CDS c. CUH ABDOMINO
MEDLINE INDUSTRIES, LP - Northfield · 2023-02-15
VTC Regular Kit Nephrostomy Catheter System Kit, Material Number REF M001245300; to provide external drainage of the uri
Boston Scientific Corporation · 2024-09-25
EnChroma Safety Glasses with Rx Indoor Universal Lenses: Martinez Cx1 Indoor Rx, SKU: Cx.PC.IN.MTZ.BK.Rx; Summit Indoo
Enchroma Inc · 2024-11-27
Otological Ventilation Tube - T-Tube 9mm - Silicone; Product Code: NZ3309;
Adept Medical Ltd · 2025-01-29